Press
【
F1
】
and
【
F2
】
at same time 3seconds
〖
P0000
〗
Second digit flash
,
use function key to
move digit, set value.
Press
【
TARE
】【
ZERO
】【
CLEAR
】
〖
P1358
〗
Input 1358
,
1358 is calibration password
Press
【
SHIFT
】
〖
SCALE
〗
Confirm
,
into calibration program
___________________________________________________________________________
___
Press
【
SHIFT
】
〖
CALSP
〗
No load
,
the screen display 0,and wait
until STB and ZERO light on.
Press
【
TARE
】【
ZERO
】【
CLEAR
】
〖
0500.0
〗
Input 500kg. wait until STB light on.
Press
【
SHIFT
】
〖
-------
〗
Waiting stable
〖
LoAd3
〗
Third load calibration
Hanging on third load
1000kg
Suppose third load is 1000kg,, wait until
STB light on.
Press
【
SHIFT
】
〖
0700.0
〗
Display third minimum load, equal to
second load plus 20% capacity, and first
digit flash.
Press
【
TARE
】【
ZERO
】【
CLEAR
】
〖
1000.0
〗
Input 1000kg. wait until STB light on.
Press
【
SHIFT
】
〖
------ -
〗
Waiting stable
〖
-oL-
〗
〖
End
〗
Calibration end
〖
000.00
〗
Display calibrated capacity , sometimes it
display 999.90,999.95,1000.05,1000.10
Press
【
TARE
】
〖
1000.0
〗
Move decimal position
Put down weight
〖
0.0
〗
Check zero point, it display 0.0, and STB,
ZERO light on.
Hanging on 1000kg
weight again
〖
1000.0
〗
Below is test program to check scale
accuracy.
Put down 1000kg weight,
hang on 10kg weight
〖
10.0
〗
10kg is minimum cap
(
20e
)
, 20kg is also
acceptable
Put down 10kg weight,
hang on 250kg weight
〖
100.0
〗
250kg is 500e, check accuracy
Put down weight
〖
0.0
〗
Check zero point
1. First load: LAOD1 must larger than 20% capacity
,
and LOAD1<LOAD2<LOAD3, at the
same time, between LOAD1 and LOAD2
,
LOAD2 and LOAD3, the interval must bigger
than 20% capacity
2. If balance of any point load is less than 20% capacity, after this point load cabliration finish,