CALIBRATION 

When to calibrate 

Calibration may be required when it is initially installed, if the scale/balance is moved 
to a substantial distance.  This is necessary because the weight of a mass in one 
location is not necessarily the same in another location. Also, with time and use, 
mechanical deviations may occur. 

How to calibrate 

1.

Before entering calibration mode, 

There is a switch key on the bottom of the 

scale. Turn on the switch key first

,  press 

[ON/OFF]

  to turn the power on for 

1minutes. 

 

2.Enter calibration mode 

Press 

[ON/OFF]

 to turn the power on. the display will show flash digits and then 

-

-------

, Press 

[ZERO]

 , the display will show “CAL” , Press 

[ZERO] 

again ,

 

the 

display will show

 “X0000 

CAL

 , 

X

 is flash digit. Press 

[UNIT] 

key the flash digit 

will move to right; Press 

[PCS]

 key to increase the flash digit.

(X is the calibration 

weight which can be setted according to your requirement)

 

3.Calibration by the weight 

When you selected the calibration weight ,then place the calibration weight on 
platform. Then press

 [ZERO]

 key

,

 the display return to the weight value, Now the 

calibration is complete. 

 

4

.Press 

[ON/OFF]

, turn off the power, press 

[ON/OFF]

 turn on the power again, 

place the weight on the platform to make sure whether weighing is correct, if not, 
repeat steps 2-4.

 

5.Return to weighing mode   

Press 

[ON/OFF]

 to turn the power off. Press 

[ON/OFF]

 again to turn on the power 

and the scale/balance return to the weighing mode. 

 

 

 

PARTS DESCRIPTION 

 

RIGHT SIDE 

     

AC adaptor 

Socket 

 

  TOP 

 

 

Display 

 

ON/OFF

 key 

Turns the scale 

 

power on/off. 

                                                  

ZERO 

key 

ets display to zero   

 

PRINT

 key 

 

UNIT 

key                             PCS key 

Changes weighing units, g ,ct, oz, tola 

 

 

 

  BOTTOM 

 

Switch key

 

Battery compartment 

 

Battery compartment cover     

 

Batteries   

Use 6 AA batteries 

 

DISPLAY   

 

Low battery indicator                  weighing unit    

TARE

 indicator 

Negative value indicator   
 

STABLE

 indicator   

when the reading is stable.     

 

 

Print indicator 

ZERO

 indicator 

 

Summary of Contents for FR-H-100

Page 1: ...ect ON activate the selected unit or OFF inactivate the selected unit press ZERO again to return to weighing mode 3 Switch the power to ON and than press ON OFF to turn on the scale The display shows F E 9 to 0 and than press PCS key to enter ZERO track function setting display shows x xd ZER press UNIT key to select ZERO track value 0 0d 0 5d 1 0d 1 5d 2 0d 3 0d press ZERO key to enter ZERO setti...

Page 2: ...Discharge the static electricity As example the method is to use the electrification prevention spray to spray it to both side of weighing platform 1 SPECIFICATIONS Model number Capacity Division FR H 100 100g 0 001g 0 005g FR H 200 200g 0 005g FR H 300 300g 0 005g 0 01g FR H 600 600g 0 01g 0 005g FR H 1000 1000g 0 01g 0 02g FR H 3000 3000g 0 05g 0 1g FR H 6000 6000g 0 1g 0 2g FR H 10000 10000g 0 ...

Page 3: ...ch can be setted according to your requirement 3 Calibration by the weight When you selected the calibration weight then place the calibration weight on platform Then press ZERO key the display return to the weight value Now the calibration is complete 4 Press ON OFF turn off the power press ON OFF turn on the power again place the weight on the platform to make sure whether weighing is correct if...

Page 4: ... accurately Error messages EEEE Overload Low battery Calibration may be required before weighing Read CALIBRATION first and if necessary calibrate your scale balance for accurate weighing Weighing procedures 1 Press ON OFF to turn on the balance When power is turned on all display segments appear on LCD for a few seconds and finally 0 will appear on the display 2 Select the weighing unit with UNIT...

Page 5: ...Bit 8 7 7 Parity None Even Odd Stop Bit 1 Code ASCII Data Format W WEIGHT M MESSAGE 1 2 3 4 5 6 7 8 9 1 0 1 1 1 2 1 3 1 4 1 5 1 6 1 7 1 8 1 9 2 0 HEA D DATA UNIT CR W M g C R L F Example 100 00g W 100 00g 500 00ct W 500 00ct Err 0 M over ...

Reviews: