CALIBRATION
When to calibrate
Calibration may be required when it is initially installed, if the scale/balance is moved
to a substantial distance. This is necessary because the weight of a mass in one
location is not necessarily the same in another location. Also, with time and use,
mechanical deviations may occur.
How to calibrate
1.
Before entering calibration mode,
There is a switch key on the bottom of the
scale. Turn on the switch key first
, press
[ON/OFF]
to turn the power on for
1minutes.
2.Enter calibration mode
Press
[ON/OFF]
to turn the power on. the display will show flash digits and then
“
-
-------
”
, Press
[ZERO]
, the display will show “CAL” , Press
[ZERO]
again ,
the
display will show
“X0000
CAL
”
,
X
is flash digit. Press
[UNIT]
key the flash digit
will move to right; Press
[PCS]
key to increase the flash digit.
(X is the calibration
weight which can be setted according to your requirement)
3.Calibration by the weight
When you selected the calibration weight ,then place the calibration weight on
platform. Then press
[ZERO]
key
,
the display return to the weight value, Now the
calibration is complete.
4
.Press
[ON/OFF]
, turn off the power, press
[ON/OFF]
turn on the power again,
place the weight on the platform to make sure whether weighing is correct, if not,
repeat steps 2-4.
5.Return to weighing mode
Press
[ON/OFF]
to turn the power off. Press
[ON/OFF]
again to turn on the power
and the scale/balance return to the weighing mode.
5
PARTS DESCRIPTION
RIGHT SIDE
AC adaptor
Socket
TOP
Display
ON/OFF
key
Turns the scale
power on/off.
ZERO
key
ets display to zero
key
UNIT
key PCS key
Changes weighing units, g ,ct, oz, tola
BOTTOM
Switch key
Battery compartment
Battery compartment cover
Batteries
Use 6 AA batteries
DISPLAY
Low battery indicator weighing unit
TARE
indicator
Negative value indicator
STABLE
indicator
when the reading is stable.
Print indicator
ZERO
indicator
2