Express V
2013-06
Page 15
2.13.3
Adjusting operator unit
L
L
L
L
CAUTION:
Risk of pinching - Do not put your fingers between the operator
control and the other parts when positioning the operator unit.
The operator unit's horizontal position can be changed:
1. Loosen screw just under the armpad a little bit.
2. Move the operator unit to the desired position or remove the operator unit.
3. Retighten screw properly.
If you pull screw to the side, the operator unit can be turned to the side.
NOTE: Depending on the features of your wheelchair, screw
could also be located
beneath the guide tube
➁
.
2.13.4
Connections for charger / program unit
L
L
L
L
WARNING:
Risk of injury - Check that all plugs (battery charger and
programming unit) are removed before setting the wheelchair in motion.
The socket for connecting the charger supplied is
located on the front of the operator unit. Here, too, is
the connnection for the programming unit, which
should only be connected and used by authorised
persons who have been trained to program it
(Vermeiren personnel).
Make sure that the joystick is in the neutral
(central) position when pressing the on/off
button, otherwise the electronics will be blocked.
This block can be lifted by switching the control
unit off and then on again.