Table 201: Guidance and manufacturer's declaration –
electromagnetic emissions
The VeriChip H2 Reader is intended for use in the electromagnetic
environment specified below. The customer or the user of the
VeriChip H2 Reader should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic
environment guidance
RF emissions
CISPR 11
Group 1
The VeriChip H2 Reader
uses RF energy only for
its internal function.
Therefore, its RF
emissions are very low
and are not likely to cause
any interference in the
nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The VeriChip H2 Reader
is suitable for use in all
establishments, including
domestic establishments
and those directly
connected to the public
low-voltage power supply
network that supplies
buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage
Fluctuations/
flicker emission
Not applicable
Table 202: Guidance and manufacturer's declaration-
electromagnetic immunity
The VeriChip H2 Reader is intended for use in the electromagnetic
environment specified below. The customer or the user of the
VeriChip H2 Reader should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic
environment
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 KV contact
±8 KV air
±6KV contact
±8 KV air
Floors should
be wood,
concrete or
ceramic tile. If
floors are
covered with
synthetic
material, the
relative humidity
should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 KV for power
supply lines
±1 KV for input/
output lines
Not applicable
Not applicable
Surge
IEC 61000-4-5
±1 KV
differential
mode
±2 KV common
mode
Not applicable
Not applicable
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5 %
U
T
(>95%
dip in
U
T
) for 0.5
cycle 40%
U
T
(60 % dip in
U
T
)
for 5 cycles 70%
U
T
(3 % dip in
U
T
)
for 25 cycles
<5%
U
T
(>95 %
dip in
U
T
) for 5
sec
Not applicable
Not applicable
Power
Frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power
frequency
magnetic fields
should be at
levels
characteristic
of a typical
location in
a typical
commercial
or hospital
environment.
Cleaning
Wipe the Reader with alcohol wipes. Do not submerge or
spray liquid directly on the Reader.