5.0 Operation
5.3
Notes on Test
The furnace is intended for use with tests in accordance with BS1016: Part107.1:1991
and ISO501:1981.
Before commencing the test, the furnace chamber lid should be opened and the
temperature allowed to stabilise for no longer than 2 minutes. The test should be
conducted with the furnace lid open. When the test is completed, close the lid to
conserve energy and prolong element life.
5.4
Operator Safety
Ceramic materials used in the furnace insulation can become conducting
at high temperatures; therefore there is a potential danger to the
operator, as this furnace is designed to be used with the chamber lid
open.
To guard against this danger, the furnace is fitted with a 30mA earth leakage (residual
current) circuit breaker mounted in the back of the case. This trips out in the event of
accidental contact with the heating element or current leakage through the insulation.
The device may be reset after tripping out. From time to time, test it using the integral
test button.