6.2.1
Cleaning
Soot deposits may form inside the furnace, depending on the process. At appropriate
intervals remove these by heating as indicated in the General Operation Notes.
The product's outer surface may be cleaned with a damp cloth. Do not
allow water to enter the interior of the case or chamber. Do not clean
with organic solvents.
6.3
Calibration
After prolonged use, the controller and/or thermocouple may require recalibration. This
is important for processes that require accurate temperature readings or for those that
use the product close to its maximum temperature. A quick check using an independent
thermocouple and temperature indicator should be made from time to time to
determine whether full calibration is required. Carbolite Gero can supply these items.
Depending on the controller fitted, the controller instructions may contain calibration
instructions.
6.4
After-Sales Service
Carbolite Gero Service has a team of Service Engineers who can offer repair, calibration
and preventive maintenance of furnace and oven products both at the Carbolite Gero
factory and at customers’ premises throughout the world. A telephone call or email
often enables a fault to be diagnosed and the necessary parts to be despatched.
In all correspondence please quote the serial number and model type given on the
rating label of the product. The serial number and model type are also given on the
back of this manual when supplied with the product.
Carbolite Gero Service and Carbolite Gero contact information can be found on the back
page of this manual.
6.5
Recommended Spare Parts and Spare Parts Kit
Carbolite Gero can supply individual spare parts or a kit of the items most likely to be
required. Ordering a kit in advance can save time in the event of a breakdown.
Each kit consists of one thermocouple, one solid state relay, one element and one lid
insulation piece.
When ordering spare kits please quote the model details as requested above.
Crucibles with lids and spare crucible holders, can also be supplied.
6.6
Power Adjustment
The control system incorporates electronic power limiting, but for the model listed in
this manual the power limit is set to 100%. The power limit parameter OP.Hi may be
accessible to the operator, but should not generally be altered.
29
6.0 Maintenance
Summary of Contents for CARBOLITE GERO SNF
Page 4: ...10 2 Power Settings 38 11 0 Specifications 39 11 1 Environment 39 4 ...
Page 20: ...3 0 2132 Controller 3 7 Navigation Diagram 20 ...
Page 28: ...6 0 Maintenance Power Consumption Measure the current drawn on each phase cir cuit 28 ...
Page 40: ......