2.0 Installation
Examples:
l
to alter a furnace made for a 240 V supply to 208 V: replace thyristor stack,
move a cable to the 208 V transformer tapping, and adjust the thyristor stack.
l
to change from a supply of 220 V to 230 V: move a cable to the 230 V, and adjust
the thyristor stack.
Please contract Carbolite Gero Service for guidance and assistance if the power supply
shown on the rating label does not match the power supply available.