Supply
Terminal Label
Cable Colour
Supply Types
Live - Neutral
Reversible or Live-
Live
1-phase L
Brown
to live
to either power
conductor
(For USA 200-240V,
connect L1)
N
Blue
to neutral
to the other power
conductor
(For USA 200-240V,
connect L2)
PE
Green/ Yellow
to earth
(ground)
to earth (ground)
Electrical Connection Details:
2-phase L1
Black
to phase 1
L2
Black
to phase 2
N
Light Blue
to neutral
PE
Green/ Yellow
to earth (ground)
2.5
Power Supply Notes
Each model is manufactured for one of the following voltage ranges:
l
208 V
l
220 to 240 V
l
380 to 415 V (not HTF 17/5 or HTF 18/4)
It is not possible to modify and use a furnace manufactured for the 380-415 range on
either of the other two voltage ranges listed above: there are too many component
differences.
It is possible to modify a furnace manufactured for either 208 and 220-240, provided
the thyristor unit is replaced. A transformer tapping must also be altered. See sections
8.4 and 8.10.
It is possible to alter the voltage within any of the ranges above, by reconnecting the
incoming cable to the transformer to the appropriate primary tapping. The tappings are
208 - 220 - 230 - 240 V, or 380 - 400 - 415 V, and are labelled on the primary side of
the transformer - see section 8.10. It is also necessary to adjust the setting of the
thyristor stack: see section 8.4.
11
2.0 Installation
Summary of Contents for CARBOLITE GERO HTF 17/10
Page 7: ...Caution Double Pole Neutral Fusing 7 1 0 Symbols and Warnings ...
Page 38: ...4 0 3508 Dual Loop Cascade Power Limit View 38 ...
Page 60: ...8 0 Repairs and Replacements 60 ...
Page 65: ...G Grey GR Y Green Yellow 65 10 0 Wiring Diagrams ...
Page 69: ...Notes Service Record Engineer Name Date Record of Work ...