6.11. Temperature Measurements 

 
Do not connect the thermocouples with electrical circuits in order to avoid electroshocks. Do not use this device when 
voltages on the measurement surface exceed 60V DC or 24V rms. AC. Do not measure temperatures in a 

microwave oven. 
1.  Put the selector in the °C position and read the current ambient temperature from the LCD. 
2.  Connect K-thermocouples with the COM and the °C terminals (or use the multifunctional socket). Observe the 

polarity! 

3.  Touch the test object with the thermocouple. 
4.  Read the stable temperature from the LCD. 
 

6.12. Current Measurements 

 
Do not perform current measurements on circuits with a potential > 600V above earth ground in order to avoid user 
injury and damage to the device. Check the fuse before you start your measurements. Use the proper terminals, 
function and range for your measurement. Never measure voltages (do not perform parallel measurements) when 

the selector is positioned on a current range. 
This device measures currents in the following ranges : 400µA, 4000µA, 40mA, 400mA, 4000A and 10A 
 
1.  Deactivate the circuit to be tested before connecting it to the meter and discharge all capacitors. 

2.  Put the selector in the µA, mA or A range. 
3.  Press the yellow button to select DCA or ACA measuring mode. 
4.  Connect the black test lead with the COM jack and the red test lead with the mA jack for measurements up to 

400mA. Move the red test lead to the A jack for measurements up to 10A. 

5.  Cut the circuit to be tested. Touch the more negative side of the break with the black test tip. Touch the more 

positive side of the break with the red test tip. Switch the two test tips in case of a negative reading. 

6.  Turn on the power to the circuit and read the display. Be sure to note the measurement units at the right side of 

the display (µA, mA or A). When only "OL" is displayed, it indicates overrange situation and the higher range has 

to be selected. 

7.  Turn off power to the circuit and discharge all capacitors. Remove the meter and restore the circuit to normal operation. 
 
 

7. Specifications 

 
Max. accuracy is achieved during a one-year period after calibration. Ideal circumstances require a temperature of 18 
to 28°C (64 to 82°F) and a max. relative humidity of 75%. 

 

7.1 Voltage 

 

Function 

Range 

Resolution 

Accuracy 

Input Impedance (nom.) 

DC mV (mV 

400mV 

0.1mV 

± 1.0% of rdg + 10 digits 

4V 1mV 

40V 10mV 

400V 100mV 

 

DC voltage (V

1000V 1V 

 

± 0.5% of rdg + 3 digits 

 
 

> 10M

 < 100pF 

AC voltage (V

400mV 

0.1mV 

± 3.0% of rdg + 3 digits 

 4V 

1mV 

 40V 

10mV 

 400V 

100mV 

 1000V 

1V 

 

± 1.0% of rdg + 3 digits 

 
 

> 5M

 < 100pF 

Overload protection : 1000V DC or AC rms 

Summary of Contents for DVM1190

Page 1: ...A connections are protected against the faulty connection of test leads The jacks that can t be used for the selected function are closed off The selector closes off all jacks that cannot be used wit...

Page 2: ...t sources before opening the meter Replace the three AAA batteries when the symbol is displayed Replace blown fuses with an identical one or a fuse with identical voltage and current ratings F 200mA 2...

Page 3: ...ng mode The measured value is saved and the selected range remains the same Cancel the data hold function by changing the measurement mode pressing the RANGE button or pressing the HOLD button again 6...

Page 4: ...CV measurement Remarks The device may display unstable readings especially in the 400mV range even though no test leads are connected with input terminals Make contact between the measuring tips of th...

Page 5: ...ctroshocks Also make sure the capacitors have been discharged fully before inserting them into the socket The device measures capacitance in the following ranges 50 00nF 500 0nF 5 000 F 50 00 F and 10...

Page 6: ...lector in the A mA or A range 3 Press the yellow button to select DCA or ACA measuring mode 4 Connect the black test lead with the COM jack and the red test lead with the mA jack for measurements up t...

Page 7: ...1k 0 5 of rdg 2 digits Resistance 40M 10k 1 5 of rdg 3 digits Overload protection 1000V DC or AC rms 7 4 Diode Test Function Range Resolution Test Condition Accuracy Diode test 1V 0 001V forward DC c...

Page 8: ...urements 5A max 4 minutes of uninterrupted measuring followed by 10 min OFF unknown above 10A 1 frequency range 40Hz 200Hz 2 average response calibrated in rms of a sine wave 8 Maintenance 8 1 General...

Page 9: ...sconnect the test leads before opening the case or the battery compartment 1 Put the selector in the OFF position 2 Disconnect all test leads and or any connectors from circuits under test 3 Open the...

Reviews: