background image

8

3 Mounting

VEGAPULS 68 • 4 … 20 mA/HART - four-wire

51905-EN-180222

1.  Clamp VEGAPULS 68 with the flange, e.g. in a bench vice
2.  Hold the connection piece (3) with a wrench (spanner size 22) on 

the flattenings

3.  Loosen counter nut (2) with spanner size 36 completely in the 

direction antenna

4.  Loosen compression nut (1) with a wrench SW 41 completely in 

the direction of the antenna

5.  Remove the parabolic antenna (4) axially
6.  Mount sensor flange on the adapter flange and fasten it tightly
7.  Check if the O-ring seal is present on the connection piece and 

make sure it is not damaged.

Note:

A damaged O-ring seal must be replaced: FKM article no. 2.28248, 

FFKM (Kalrez 6375) article no. 2.27351
8.  Remount the parabolic antenna (4)
9.  Tighten compression nut (1) with SW 41, torque max. 50 Nm
10.  Tighten counter nut (2) with SW 36, torque max. 40 Nm.

Note:

Take note for VEGAPULS 68 with rinsing air connection that the holes 

in the antenna and in the process fitting correspond. This ensures a 

sufficient air flow (the air is led through the holes to the feed system. A 

rinsing of the parabolic antenna in total is not intended).

1

2
3
4

Fig. 2: Dismounting, parabolic antenna
1  Compression nut

2  Counter nut

3  Connection piece

4  Parabolic antenna

3.3  Mounting instructions

1.  Distance from the vessel wall > 200 mm, the antenna should 

protrude > 10 mm into the vessel

Mounting

Summary of Contents for VEGAPULS 68

Page 1: ...Quick setup guide Radar sensor for continuous level measurement of bulk solids VEGAPULS 68 4 20 mA HART four wire Document ID 51905...

Page 2: ...ting instructions 8 4 Connecting to power supply 10 4 1 Connection procedure 10 4 2 Wiring plan double chamber housing 11 5 Set up with the display and adjustment module PLICSCOM 12 5 1 Insert display...

Page 3: ...em components through incorrect mounting or adjustment Thus dam age to property to persons or environmental contamination can be caused Also the protective characteristics of the instrument can be inf...

Page 4: ...netic interference from electrical conductors and radiated emissions must be taken into account as is usual with a class A instrument according to EN 61326 1 If the instrument is used in a different e...

Page 5: ...Take note of the respective operating regulations This device may not cause interference and The device must be resistant to interference signals including such that may cause undesired operating stat...

Page 6: ...orks With the serial number you can access the delivery data of the instru ment via www vega com VEGA Tools and Instrument search You can find the serial number on the inside of the instrument as well...

Page 7: ...it against falling off 4 Retighten the antenna with hexagon screws to the antenna socket torque max 10 Nm 7 5 lbf ft Note The VEGAPULS 68 with rinsing air connection or antenna exten sion is provided...

Page 8: ...maged Note A damaged O ring seal must be replaced FKM article no 2 28248 FFKM Kalrez 6375 article no 2 27351 8 Remount the parabolic antenna 4 9 Tighten compression nut 1 with SW 41 torque max 50 Nm 1...

Page 9: ...ire 51905 EN 180222 ca 10 mm 200 mm 7 87 Fig 3 Distance of the antenna to the vessel wall vessel ceiling 2 Note min socket diameter depending on the socket length 3 Note the instructions for sealing F...

Page 10: ...sensor through the cable entry 6 Lift the opening levers of the terminals with a screwdriver see following illustration 7 Insert the wire ends into the open terminals according to the wir ing plan Fi...

Page 11: ...51905 EN 180222 4 2 Wiring plan double chamber housing The following illustrations apply to the non Ex as well as to the Ex ia version 4 20 mA PE L N L1 N GND 1 2 1 2 3 4 4 20mA IS Fig 5 Wiring plan d...

Page 12: ...on the electronics you can choose any one of four different posi tions each displaced by 90 3 Press the display and adjustment module onto the electronics and turn it to the right until it snaps in 4...

Page 13: ...000 m d 4 000 m d 3 Select in the menu item Medium the medium of your applica tion for example Powder Dust Medium Bulk solid 4 Select in the menu item Application the vessel and the vessel form for e...

Page 14: ...ng the product level These settings can be made ahead of time without the instrument having to be installed High sockets or vessel installations such as e g struts or agitators as well as buildup and...

Page 15: ...in this area 5 3 Menu schematic Information Depending on the version and application the light coloured menu windows are not always available or offer nor selection possibility Basic adjustment Basic...

Page 16: ...play Diagnostics Service Info False signal suppression 4 1 Change now Extended setting 4 2 None Quick level change Current output 4 3 Characteristic 4 20 mA Fail mode 3 6 mA min current 4 mA max curre...

Page 17: ...supply Electromechanical data version IP 66 IP 67 Cable entry Double chamber housing 1 x cable gland M20 x 1 5 cable 5 9 mm 1 x blind plug M20 x 1 5 plug M12 x 1 for VEGADIS 61 optional or 1 x closin...

Page 18: ...18 Notes VEGAPULS 68 4 20 mA HART four wire 51905 EN 180222...

Page 19: ...19 Notes VEGAPULS 68 4 20 mA HART four wire 51905 EN 180222...

Page 20: ...of delivery application practical use and operat ing conditions of the sensors and processing systems correspond to the information available at the time of printing Subject to change without prior n...

Reviews: