C
ycle time measuring cycle approx
.
-
S
tandard electronics
0
.
5
s
-
E
lectronics with increased sensitivity
1
s
S
ignal resolution
0
.
3
µ
A
R
esidual ripple
±
0
.
4
µ
A
F
ailure signal current output
(
adjustable
)
m
A
-
value unchanged
20
.
5
m
A
,
22
m
A
,
<
3
.
6
m
A
M
ax
.
output current
22
m
A
S
tarting current
≤
3
.
6
m
A
;
≤
10
m
A
for
5
ms after switching on
L
oad
see load diagram under
P
ower supply
D
amping
(
63
%
of the input variable
)
0
…
999
s
,
adjustable
HART
output values according to
HART
7
.
0
1
)
-
PV
(
P
rimary
V
alue
)
D
istance to the level
-
SV
(
S
econdary
V
alue
)
L
evel as percentage value
-
TV
(
T
hird
V
alue
)
L
inearised percentage value
-
QV
(
F
ourth
V
alue
)
S
caled measured value
R
esolution
,
digital
<
1
mm
(
0
.
039
in
)
A
ccuracy
(
similar to
DIN EN
60770
-
1
)
P
rocess reference conditions according to
DIN EN
61298
-
1
-
T
emperature
+
18
…
+
30
°
C
(
+
64
…
+
86
°
F
)
-
R
elative humidity
45
…
75
%
-
A
ir pressure
860
…
1060
mbar
/
86
…
106
k
P
a
(
12
.
5
…
15
.
4
psig
)
I
nstallation reference conditions
-
M
in
.
distance to installations
>
200
mm
(
7
.
874
in
)
-
R
e
fl
ector
C
orner re
fl
ector
-
F
alse re
fl
ections
L
argest false echo
20
d
B
smaller than the useful
echo
D
eviation with liquids
S
ee following diagrams
1
)
D
efault values
,
can be assigned individually
56
VEGAPULS
62
•
4
…
20
m
A
/
HART
two
-
wire
11
S
upplement
36503
-
EN
-
100607