E
nter the requested parameters via the appropriate keys
,
save your
settings and jump to the next menu item with the
[
ESC
]
and
[-
>
]
key
.
C
aution
:
I
f the instrument is used as part of an over
fi
ll protection system
according to
WHG
(
W
ater
R
esources
A
ct
),
please note the following
:
I
f a linearisation curve is selected
,
the measuring signal is no longer
compulsorily linear proportional to the level
.
T
his must be taken into
consideration by the user
,
particularly when adjusting the switching
point on the level switch
.
W
ith radar sensors with standpipe antenna
,
the sensor length is
already preset in this menu item
.
W
hen shortening the standpipe
antenna afterwards
,
this value must be corrected respectively
.
W
ith a reset
,
all settings despite a few exceptions will be reset
.
T
he
exceptions are
:
PIN
,
language
,
lighting
,
SIL
and
HART
mode
.
T
he following reset functions are available
:
D
elivery status
:
R
estoring the parameter settings at the time of
shipment from the factory incl
.
the order
-
speci
fi
c settings
.
A
created
false signal suppression
,
user
-
programmable linearization curve as
well as the measured value memory will be deleted
.
B
asic settings
:
R
esetting the parameter settings incl
.
special and
laboratory parameters to the default values of the respective instru
-
ment
.
A
created false signal suppression
,
user programmable
linearization curve as well as the measured value memory will be
deleted
.
A
dditonal settings
/
S
en
-
sor length
A
dditional settings
-
R
e
-
set
VEGAPULS
62
•
4
…
20
m
A
/
HART
two
-
wire