C
ode
T
ext mes
-
sage
C
ause
R
ecti
fi
cation
PA D
ev
-
S
pec
D
iagnosis
F
105
D
etermine
measured
value
l
T
he instrument is still
in the start phase
,
the
measured value
could not yet be de
-
termined
l
W
ait for the warm
-
up
phase
l
D
uration depending
on the version and
parameter adjust
-
ment up to approxi
-
mately
3
min
.
B
it
6
F
113
C
ommuni
-
cation error
l
E
rror in the internal
instrument communi
-
cation
l
S
eparate operating
voltage brie
fl
y
l
S
end instrument for
repair
B
it
7
F
125
U
npermis
-
sible elect
-
ronics tem
-
perature
l
T
emperature of the
electronics in the
non
-
speci
fi
ed section
l
C
heck ambient tem
-
perature
l
I
solate electronics
l
U
se instrument with
higher temperature
range
B
it
8
F
260
E
rror in the
calibration
l
E
rror in the calibra
-
tion carried out in the
factory
l
E
rror in the
EEPROM
l
E
xchange of the
electronics
l
S
end instrument for
repair
B
it
9
F
261
E
rror in the
con
fi
gurati
-
on
l
E
rror during setup
l
F
alse signal sup
-
pression faulty
l
E
rror when carrying
out a reset
l
R
epeat setup
l
R
epeat reset
B
it
10
F
264
I
nstallation
/
S
etup error
l
A
djustment not within
the vessel height
/
measuring range
l
M
ax
.
measuring ran
-
ge of the instrument
not su
ffi
cient
l
C
heck or correct in
-
stallation and
/
or pa
-
rameter adjustment
l
U
se an instrument
with bigger measu
-
ring range
B
it
11
F
265
M
easure
-
ment func
-
tion distur
-
bed
l
S
ensor does no lon
-
ger carry out a mea
-
surement
l
O
perating voltage too
low
l
C
heck operating vol
-
tage
l
C
arry out a reset
l
S
eparate operating
voltage brie
fl
y
B
it
12
T
he following table shows the error codes and text messages in the
status message
"
F
unction check
"
and provides information on causes
as well as corrective measures
.
F
unction check
38
VEGAPULS
62
standpipe version
•
F
oundation
F
ieldbus
9
D
iagnosis
,
A
sset
M
anagement and service
36510
-
EN
-