M
aintenance
:
D
ue to external in
fl
uences
,
the instrument function is
limited
.
T
he measurement is a
ff
ected
,
but the measured value is still
valid
.
P
lan in maintenance for the instrument because a failure is
expected in the near future
(
e
.
g
.
due to buildup
).
T
his status message is inactive by default
.
A
ctivation by the user via
PACT
ware
/
DTM
or
EDD
is possible
.
T
he following table shows the codes and text messages of the status
message
"
F
ailure
"
and provides information on causes as well as
corrective measures
.
C
ode
T
ext mes
-
sage
C
ause
R
ecti
fi
cation
PA D
ev
-
S
pec
D
iagnosis
F
013
no measu
-
red value
available
l
S
ensor does not de
-
tect an echo during
operation
l
A
ntenna system con
-
taminated or defecti
-
ve
l
C
heck or correct in
-
stallation and
/
or pa
-
rameter adjustment
l
C
lean or exchange
process component
or antenna
B
it
0
F
017
A
djustment
span too
small
l
A
djustment not within
speci
fi
cation
l
C
hange adjustment
according to the limit
values
(
di
ff
erence
between min
.
and
max
.
≥
10
mm
)
B
it
1
F
025
E
rror in the
linearization
table
l
I
ndex markers are not
continuously rising
,
for examle unlogical
value pairs
l
C
heck linearization
table
l
D
elete table
/
C
reate
new
B
it
2
F
036
N
o operab
-
le software
l
F
ailed or interrupted
software update
l
R
epeat software up
-
date
l
C
heck electronics
version
l
E
xchange of the
electronics
l
S
end instrument for
repair
B
it
3
F
040
E
rror in the
electronics
l
H
ardware defect
l
E
xchange of the
electronics
l
S
end instrument for
repair
B
it
4
F
080
l
G
eneral software er
-
ror
l
S
eparate operating
voltage brie
fl
y
B
it
5
F
ailure
(
failure
)
VEGAPULS
62
standpipe version
•
F
oundation
F
ieldbus
37
9
D
iagnosis
,
A
sset
M
anagement and service
36510
-
EN
-
120301