2
1
F
ig
.
9
:
F
ront
fl
ush installation
1
U
nfavourable installation
-
bulk solid can settle in front of the sensor
2
C
orrect front
fl
ush installation
I
f a front
-
fl
ush installation is not possible
,
you can
fi
ll long sockets with
a plastic or ceramic cylinder to avoid build up of bulk solids and dust
deposits
.
S
elect the material according to the diameter of the socket
and adapt the length as far as possible to the shape of the vessel wall
.
S
ecure the
fi
lling material against falling out or turning
.
I
f using a
fi
screw is technically not possible
,
you can secure the material in the
socket with an adhesive
.
18
VEGAMIP R
61
• -
T
ransistor
4
M
ounting
40694
-
EN
-
120228