32
6 Set up with the display and adjustment module PLICSCOM
VEGACAL 62 • Profibus PA
30025-EN-160331
Backlight
In the default setting, the lightning is switched off.
The respective min. and max. measured values are saved in the sen-
sor. The values are displayed in the menu item "
Peak values
".
•
Min. and max. distance in m(d)
•
Min. and max. temperature
Peak value indicator
The instrument status is displayed in this menu item. If no failure is
detected by the sensor, "
OK
" will be displayed. If a failure is detected,
there will be a sensor-specific flashing fault signal, for example
"
E013
". The failure is also displayed in clear text, for example "
No
measured value available
".
Information:
The fault message as well as the clear text indication are also carried
out in the measured value display.
Meas. certainty
Sensor status
Up to 3000 measured values are recorded (depending on the sensor)
when starting a "
Trend curve
". Then the values can be displayed on
a time axis. The oldest measured values are always deleted.
The measured values displayed are in the unit pF.
Trend recording
Presentation of the trend
curve
Information:
The trend recording is not activated when being shipped. It must be
started by the user via the menu item "
Start trend curve
".
Profibus transmits two values cyclically. The first value is determined
in the menu item "
Channel
". The selection of the additional cyclical
value is made in the menu item "
Additional PA value
".
The following values are available with radar, guided microwave and
ultrasonic sensors:
Diagnosis - Peak value
Diagnostics - Device
status
Trend recording
Service - Additional PA
value