71
11 Supplement
VEGABAR 82 • Profibus PA
45031-EN-220624
SIP process temperature
(SIP = Sterilization in place)
Applies to instruments configurations suitable for vapour, i.e. material measuring cell seal EPDM or
FFKM (Perlast G74S).
Vapour stratification up to 2 h
+150 °C (+302 °F)
Process pressure
Permissible process pressure
see specification "
process pressure
" on the type label
Mechanical stress
11)
Vibration resistance
4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Shock resistance
50 g, 2.3 ms according to EN 60068-2-27 (mechanical
shock)
12)
Electromechanical data - version IP66/IP67 and IP66/IP68 (0.2 bar)
13)
Options of the cable entry
Ʋ Cable entry
M20 x 1.5; ½ NPT
Ʋ Cable gland
M20 x 1.5, ½ NPT (cable ø see below table)
Ʋ Blind plug
M20 x 1.5; ½ NPT
Ʋ Closing cap
½ NPT
Material cable gland/Seal insert
Cable diameter
5 … 9 mm
6 … 12 mm
7 … 12 mm
10 … 14 mm
PA/NBR
●
●
–
●
Brass, nickel-plated/NBR
●
●
–
–
Stainless steel/NBR
–
–
●
–
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, stranded wire
0.2 … 2.5 mm² (AWG 24 … 14)
Ʋ Stranded wire with end sleeve
0.2 … 1.5 mm² (AWG 24 … 16)
Electromechanical data - version IP66/IP68 (1 bar)
Connection cable, mechanical data
Ʋ Configuration
Wires, strain relief, breather capillaries, screen braiding,
metal foil, mantle
Ʋ Standard length
5 m (16.4 ft)
Ʋ Min. bending radius (at 25 °C/77 °F)
25 mm (0.984 in)
Ʋ Diameter
approx. 8 mm (0.315 in)
Ʋ Colour - version PE
Black
Ʋ Colour - version PUR
Blue
Connection cable, electrical data
Ʋ Wire cross-section
0.5 mm² (AWG 20)
11)
Depending on the instrument version.
12)
2 g with housing version stainless steel double chamber
13)
IP66/IP68 (0.2 bar), only with absolute pressure.