6 Set up with the display and adjustment module
VEGABAR 81 • 4
… 20 mA
45025-EN-140228
With a reset, certain parameter adjustments carried out by the user
are reset.
The following reset functions are available:
Delivery status:
Restoring the parameter settings at the time of
shipment from the factory incl. the order-specific settings. A user-pro
-
grammable linearization curve as well as the measured value memory
will be deleted.
Basic settings:
Resetting the parameter settings incl. special
parameters to the default values of the respective instrument. A user
programmable linearization curve as well as the measured value
memory is deleted.
The following table shows the default values of the instrument. De
-
pending on the instrument version or application, all menu items may
not be available or some may be differently assigned:
Reset - Setup
Menu item
Parameter
Default value
Measurement
loop name
Sensor
Application
Application pro
-
cess pressure
Application level
No reset
Units
Unit of measure-
ment
mbar (with nominal measuring range
≤400 mbar)
bar (with nominal measuring rang
-
es ≥1 bar)
Temperature unit °C
Position correc-
tion
0.00 bar
Adjustment
Zero/Min. adjust-
ment
0.00 bar
0.00 %
Span/Max. adjust-
ment
Nominal measuring range in bar
100.00 %
Damping
Integration time
0.0 s
Additional adjustments
- Reset