![Vega MINITRAC 31 Operating Instructions Manual Download Page 75](http://html1.mh-extra.com/html/vega/minitrac-31/minitrac-31_operating-instructions-manual_3818841075.webp)
75
11 Supplement
MINITRAC 31 • Foundation Fieldbus
41782-EN-211203
Ʋ Internal load
250 Ω
Ʋ Input voltage
max. 6 V
Switching input
Ʋ Type of input - Open Collector
10 mA
Ʋ Type of input - Relay contact
100 mA
Ʋ Input voltage
max. 24 V
Output variable - Level measurement
Output
Ʋ Signal
digital output signal, Foundation Fieldbus protocol
Ʋ Physical layer
according to IEC 61158-2
Damping (63 % of the input variable)
0 … 999 s, adjustable
Channel Numbers
Ʋ Channel 1
Process value (level)
Ʋ Channel 8
Electronics temperature
Ʋ Channel 9
Pulse rate
Transmission rate
31.25 Kbit/s
Current value
10 mA, ±0.5 mA
Resolution, digital
> 0.1 mm (0.004 in)
Output variable - Density measurement
Output
Ʋ Signal
digital output signal, Foundation Fieldbus protocol
Ʋ Physical layer
according to IEC 61158-2
Damping (63 % of the input variable)
0 … 999 s, adjustable
Channel Numbers
Ʋ Channel 1
Process value (density)
Ʋ Channel 3
Process temperature (with external temperature sensor)
Ʋ Channel 8
Electronics temperature
Ʋ Channel 9
Pulse rate
Transmission rate
31.25 Kbit/s
Current value
10 mA, ±0.5 mA
Resolution, digital
> 0.1 mm (0.004 in)
Output variable - Level detection
Output
Ʋ Signal
digital output signal, Foundation Fieldbus protocol
Ʋ Physical layer
according to IEC 61158-2
Damping (63 % of the input variable)
0 … 999 s, adjustable
Channel Numbers
Ʋ Channel 1
Process value (switching status)
Ʋ Channel 8
Electronics temperature