6 Set up with the display and adjustment module
MINITRAC 31 • Foundation Fieldbus
41782-EN-211203
When a reset is carried out, all settings (with only a few exceptions)
are reset. The exceptions are: PIN, language, SIL and HART mode.
The following reset functions are available:
Basic settings:
Resetting of the parameter adjustments to default
values at the time of shipment. Order-specific settings are deleted.
Default settings:
Resetting of the parameter adjustment like under
"
Basic settings
". In addition, special parameters are reset to default
values. Order-specific settings are deleted.
Peak values of measured value:
Resetting of the parameter adjust-
ments in the menu item "
Setup
" to the default values of the respec-
tive instrument. Order-specific settings remain but are not taken over
into the current parameters.
Peak values of temperature:
Resetting of the measured min. and
max. temperatures to the actual measured value.
The following table shows the default values of the instrument. The
values apply for the application "
Density
". First of all you have to
select the application.
Depending on the instrument version, not all menu items may be
available or they may be differently assigned:
Menu
Menu item
Default value
Setup
Measurement loop name
Sensor
Isotope
Cs-137
Application
Density
Background radiation
0 ct/s
Process value unit
g/cm
3
Temperature unit
° C
Adjustment min. process value
0.500 g/cm
3
Adjustment max. process value
1.500 g/cm
3
Inner diameter
0.20 m
Linearisation
90000 ct/s and 0.500 g/cm
3
Damping
60 s
Real value correction
0
Reference value - Relay
None
Lock adjustment
Released
Reset