6 Set up with the display and adjustment module
MINITRAC 31 • Foundation Fieldbus
41782-EN-211203
To increase the Delta I value, you have to increase the distance of the
min. and max. adjustment points in the linearization.
In this menu item you can simulate measured values via the signal
output. This allows the signal path to be tested, e.g. through down-
stream indicating instruments or the input card of the control system.
Note:
To carry out a simulation with the display and adjustment module,
you have to switch on the simulation switch on the electronics module
(switch position 1).
You can find the rotary switch on the electronics module in the elec-
tronics and connection department (large cover).
You can simulate different values:
Pulse rate of the sensor
Process value
Switching function of the relay
Information:
The simulation is automatically terminated 10 minutes after the last
pressing of a key.
6.4.4 Additional adjustments
In this menu item you can set the actual date, time and display format.
Simulation
Date/Time