Maintenance and Diagnostics
Sensor Voltage
Status - Pass/Fail
Sensor high voltage fail/High voltage
on PMT is outside usable range.
Contact VEGA Field Service.
Standardize Due -
No/Yes
Standardize Due
Perform a new standardization
procedure.
Shutter Check Due
- No/Yes
Shutter Check Due
Perform shutter check. Acknowledge
on
Source Functions
tab.
New Hardware
Found - No/Yes
New hardware found?/CPU board
detects configuration mismatch. CPU
board or sensor assembly may have
been replaced or one of EEPROM
configurations is incorrect.
Contact VEGA Field Service first. If
they concur, click
Diagnostics
,
New
Hardware
,
No New Hardware
and
click
OK
.
Process Out of
Range - No/Yes
Process out of measurement
range/Current process value is not
within limits set by
Max
level and
Min
level in gauge span settings.
Contact VEGA Field Service.
X-Ray Alarm -
No/Yes
Note that there are high levels of x-ray
radiation in area that can affect
process measurement.
Contact VEGA for more information.