Setup and Calibration
The accuracy of the calibration depends upon the accuracy of the sample.
Obtain an accurate reading, from a laboratory, of the sample read by the
gauge.
Two Point Process Calibration (Preferred Method)
You must provide the high and low process conditions when performing a two-point
calibration. If using the standard method, record both the sensor counts and levels at
each step:
Set Low Density
1.
Click
Calibration
,
2 Point Calibration
,
Cal Low Collect
.
2.
Click
Start
.
3.
Click
Accept
.
4.
Enter the actual process value (from the laboratory) in engineering units.
5.
Click
OK
.
Set High Density
1.
Click
Calibration
,
2 Point Calibration
,
Cal High Collect
.
2.
Click
Start
.
3.
Click
Accept
.
4.
Enter the actual process value (from the laboratory) in engineering units.
5.
Click
OK
.
Collect Linearizer Table Data
1.
Click
Calibration
,
Linearizer Data Pt
,
Create Data Point
.
2.
At the prompt, enter the known process value.
3.
Accept or reject the results when they appear.
4.
Repeat the procedure for all available values
Calculate Calibration
1.
Click
Calibration
,
2 Point Calibration
.
2.
Click
Calculate Results
.
3.
Click
OK
.
4.
Click
OK
.