9. Press “MODE” at the air pressure calibrate mode, the device enters the
ramp time setting mode. The LCD displays “Delay Time xx min”, xx is
ramp time. Use “+” or “-” to change this value. The ramp time adjusting
range is 10 to 60.
10. Press “MODE” at the ramp time setting mode, the device enters the
starting air pressure setting mode. The LCD displays “Ramp Start
Pressure xx.x cmH2o”, xx.x is the numeric value of the starting air
pressure. Use “+”and “-”to change this value
— —
the value will
increase/reduce 0.5 once with a single press and will take a fast change
with a speed of 0.5 with continues press.
In 6,7,8,9,10, symbol “
Setting
” will flash at the frequency of 1HZ.