R404-1 

11. Press “MODE” at the starting air pressure setting mode, the device enters 

the 

compliance time

 checking mode. The LCD displays “Compliance 

Hrs xxxx.x”, xxxx.x is the compliance time of the device. This value is not 
to be changed, but only to be reset. Press “+” + “-” for a period of time of 
2 seconds to reset it.   

 

12. Press “MODE” at the compliance time checking mode, the device enters 

the 

operation time

 checking mode. The LCD displays “Operation Hrs 

xxxx.x”, xxxx.x is the operation time of the device. This value is not to be 
changed, but only to be reset. Press “+” + “-” for a period of time of 2 
seconds to reset it.   

 

13. Press “MODE” at the operation time checking mode, the device enters the 

total 

operation time

 checking mode. The LCD displays “Unit Hrs xxxx.x”, 

xxxx.x is the total operation time of the device after leaving factory. This 
value is not to be changed, but only to be reset. Press “+” “-” “+” “-” “+” “-” 
within 2 seconds to reset it. (This combination should not appear in the 
user’s manual) 

Summary of Contents for CP-03

Page 1: ...1 INSTRUCTION VEGA MANUAL Continuous Positive Airway Pressure MODEL CP 03 READ THIS INSTRUCTION Version 1 0 Date Nov 04 2009 ISO9001 ISO13485 Certified This device complies with EU EMC standard EN6060...

Page 2: ...device uses a motor which drives a fan to create a continuous air pressure to force the air into the patient s lung through the nasal cavity so that the patient s airway is expedite which is the key t...

Page 3: ...R404 1 B Assemblies...

Page 4: ...cts are being used 4 Disconnect the unit from the electrical outlet before cleaning filling and after each use 5 Do not use other attachments unless recommended by manufacturer Operating Cautions 1 Co...

Page 5: ...upply 12V DC 2 Current Consuming Max 1 5A 3 Air Pressure Range 3 cmH2O 20 cmH2O 4 Ramp Time 0 60 min 5 Air Pressure Calibrate 3 cmH2O 6 Working Flow 3 0 3m 3 h 9 cmH2O 7 Noise 33 dB 9 cmH2O 1m away 8...

Page 6: ...nd the device returns to the standby mode 4 Press RAMP at the normal working mode CPAP runs at the ramp mode the LCD outputs RAMP and Pressure xx x cmH2o xx x is the numeric value of the air pressure...

Page 7: ...ing on the screen the LCD displays Mask Off meanwhile the backlight flashes at the frequency of 1 Hz for 1 minute The following are configuration operations 5 Press RAMP and at the same time in standb...

Page 8: ...feet 2500 5000 feet 5000 7500 feet and 7500 9000 feet respectively Set this with and according to the altitude of your current location 8 Press MODE at the altitude setting mode the device enters the...

Page 9: ...Press MODE at the ramp time setting mode the device enters the starting air pressure setting mode The LCD displays Ramp Start Pressure xx x cmH2o xx x is the numeric value of the starting air pressur...

Page 10: ...he operation time checking mode The LCD displays Operation Hrs xxxx x xxxx x is the operation time of the device This value is not to be changed but only to be reset Press for a period of time of 2 se...

Page 11: ...on the screen Disconnect the USB cable the device goes back to Standby mode VI Cleaning and Sterilization It is recommended that the air tube and mask be thoroughly cleaned with hot water after each u...

Page 12: ...Submerge these parts in the vinegar and water solution for thirty minutes 5 Complete Rinsing Steps 3 5 Filter change 1 Do not wash or clean the filter And do not operate without a filter 2 Change the...

Page 13: ...R404 1 Consult accompanying documents Type BF equipment Disposal for separate collection Device in protection class 2...

Reviews: