R404-1
2. Press “START/OFF” at the standby mode, then the CPAP runs with an
output of pre-set air pressure, meanwhile, the LCD displays “
Pressure
xx.x cmH2o”(xx.x is the numeric value of the pre-set air pressure). This is
the normal working mode of the device.
3. Press “START/OFF” at the normal working mode, then the air pressure
outputting stops, and the device returns to the standby mode.
4. Press “RAMP” at the normal working mode, CPAP runs at the ramp mode,
the LCD outputs “RAMP” and “
Pressure
xx.x cmH2o”(xx.x is the numeric
value of the air pressure), the device first runs with the start pressure, 5
minutes before the pre-set delay time ends, the air pressure begins to
increase until it reaches the pre-set value at the end of the pre-set delay
time, then the device runs at the normal working mode with the pre-set
air pressure. Use the procedure described in clause 3 to stop the device.