RS-232 Commands
Function Code: 50E
34
Function Code: 50E
Function Type:
Set Temperature Compensation Value
Version 1
Command Format:
Set:
Inquire:
Display:
<SOH>S50E00TTT.T
Display:
<SOH>I50E00
Computer:
<SOH>s50E00FFFFFFFF
Computer:
<SOH>i50E00
Notes:
1.
TTT.T -
Compensation Temperature, Degrees and tenths (Decimal)
2.
FFFFFFFF -
Compensation Temperature, Degrees (ASCII Hex IEEE float)
Typical Response Message, Display Format:
<SOH>
I50E00
TTT.T
<ETX>
Typical Response Message, Computer Format:
<SOH>i50E000000000000FFFFFFFF&&CCCC<ETX>
Notes:
1.
0000000000 -
Not Used.
2.
FFFFFFFF -
Compensation Temperature, Degrees (ASCII Hex IEEE float)
3.
&& -
Data Termination Flag
4.
CCCC -
Message Checksum