C&C Mode
C&C Mode Setup Categories
45
MultiCalibrate
Figure 43 illustrates meter MultiCalibration setup procedure.
Figure 43. Meter MultiCalibration Setup
Press ENTER to accept
entry and return to
MULTICALIBRATE
START DISPENSE
DISPENSING
START
ENTER NTC VOL
FINISH
CALB RATE 1
CALB RATE 8
+
-
+
Note: In this multipoint calibration option you dispense at
up to 8 different flow rates. These rates are determined by
how much you squeeze the nozzle handle during each
calibration.
For example, the first rate you dispense product at a very
slow rate, but for each of the additional calibration rates
you increase the flow. Once calibration is complete, the
EMR3 will calculate the amount of product dispensed
based on interpolation.
Note: Dispense product into a proving
can for the calibration rate. While
dispensing, the flow rate will appear in
the Register Field and the Volume will
appear in the Preset Field. Stop when
you have dispensed a calibrated
amount.
Note: read/observe proving can
volume for this calibration rate
and enter that volume here. Press
up/down arrows to scroll through
the numbers, and NEXT to move
from one digit to the next.
Empty proving can and repeat this
procedure for Calb Rate 2. Continue until
the desired number of Calibration Rates
are complete. When finished the desired
number press ENTER until you return to
METER CALIBRTN.
RETURN TO
PROD CAL 1
(FIGURE 41)
(FIGURE 41)