Vectra Genisys® Therapy System
SAFETY PrECAuTiONS
5
Stimulation should not be applied over the anterior neck or mouth .
•
Severe spasm of the laryngeal and pharyngeal muscles may occur
and the contractions may be strong enough to close the airway or
cause difficulty in breathing .
Stimulation should not be applied transthoracically in that the
•
introduction of electrical current into the heart may cause cardiac
arrhythmia .
Stimulation should not be applied over swollen, infected, and
•
inflamed areas or skin eruptions, e .g ., phlebitis, thrombophlebitis,
varicose veins, etc .
Stimulation should not be applied over, or in proximity to, cancerous
•
lesions .
Output current density is related to electrode size . Improper
•
application may result in patient injury . If any question arises as to
the proper electrode size, consult a licensed practitioner prior to
therapy session .
The Vectra Genisys Therapy System optional modules and associated
•
accessories are designed for use only with the Chattanooga Vectra
Genisys Electrotherapy and Combination Therapy Systems .
Remove the Ultrasound or Laser Applicator by pulling the cable
•
connector only . DO NOT remove by pulling the cable .