![Vectra Fitness Genisys 27079 User Manual Download Page 112](http://html1.mh-extra.com/html/vectra-fitness/genisys-27079/genisys-27079_user-manual_859153112.webp)
Vectra Genisys® Therapy System
103
OPErATiON
NOTE:
Modality parameters may
be edited at any time during
the therapy session . Refer to
page 48
for Electrotherapy and
page 52
for Ultrasound .
CLiNiCAL rESOurCES LiBrArY- uSiNg uSEr PrOTOCOLS (CONTiNuEd)
Give Patient Interrupt Switch to patient
and explain that pressing the Red button
once pauses the therapy session .
Set intensity by rotating the Intensity Control
Knob to the prescribed level .
Set Modality Intensity
PrESSInG buTTOn OncE PauSES SESSIOn
InTEnSITy
DISPlayED
rOTaTE
InTEnSITy
knOb
Clockwise-
Increases Intensity
Counterclockwise-
Decreases Intensity
Intensity Knob Rotation
If Patient Interrupt Switch is depressed, the
treatment will be paused and a message
will appear on the System screen .
Press any button to clear the message .
NOTE:
If the Patient Interrupt Switch
is depressed a second time, the
message will clear from the screen
and the treatment will remain
paused .
Start Treatment
Press the Start button to begin therapy
session .
STarT
buTTOn