4.0   OPERATION 

4.1   INTENDED USE 

The furnace is a general purpose laboratory and heat treating of 
material and should not be used for anything else other than its 
intended purpose. 

4.2   The instructions for operating the controller, over temperature 
controller if fitted and timer if fitted are given in separate booklets. 

4.3   PRINCIPLES OF OPERATION 

The furnace is heated by a wire element wrapped around a ceramic 
muffle which is cemented into place. The muffle is placed in a box 
and insulated with a mixture of ceramic fibre and board for models 
1 –4 and wire elements suspended on ceramic tubes for models 5 & 
6. The temperature is controlled by a pid controller using a type r 
thermocouple to control the temperature. A door safety switch is 
fitted to remove the power to the heating element when the door 
is opened. Mains on the green light will illuminate when power is 
applied to the furnace. Instruments on / off switch and the 
controller will illuminate when turned to the on position. Heater on 

The amber light will illuminate whenever the elements are drawing 
power. Over temperature optional extra. The door safety switch 
disconnects the power from the elements when the door is 
opened. The amber heater on light will go out while the door is 
open if this does not happen consult the trouble shooting guide. 

Summary of Contents for MRF Series

Page 1: ...BOVE VECSTAR EQUIPMENT AND SHOULD BE READ AND UNDERSTOOD BEFORE INSTALLATION AND OPERATION CONTENTS SECTION 1 0 SYMBOLS AND WARNING LABELS 2 0 MANUAL HANDLING UNPACKING 3 0 INSTALLATION SAFETY 4 0 OPE...

Page 2: ...furnace may be hot Earth bonding point for electrical safety 1 2 LIGHTS SWITCHES Green mains on light illuminates when the supply is connected to the furnace Green instrument switch illuminates when t...

Page 3: ...taken not to trap fingers under the feet 2 2 UNPACKING Visually check for any damage to the shipping container Visually check for any damage to the equipment Open the furnace door and remove any packa...

Page 4: ...the furnace to allow adequate ventilation Ensure the furnace is placed in a way it can easily be turned off or disconnected from the electrical supply If the furnace is to be used to heat materials t...

Page 5: ...g electrical supplies must be fused correctly If a cable is fitted either wire straight to a fused isolator or fit a fused line plug Please Note The supply cable connections are Live Number 1 Neutral...

Page 6: ...bjects To avoid electrical shock this furnace must 1 Use a properly grounded electrical outlet of correct voltage and current handling capacity 2 Be disconnected from the power supply prior to mainten...

Page 7: ...wire elements suspended on ceramic tubes for models 5 6 The temperature is controlled by a pid controller using a type r thermocouple to control the temperature A door safety switch is fitted to remo...

Page 8: ...oad is to be heated uniformly it should not occupy more than two thirds of the furnace chamber If you are heating a number of small parts spread them throughout the centre two thirds of the furnace ch...

Page 9: ...tted this must be set to the correct temperature before starting a heat cycle 5 0 MAINTAINANCE 5 1 CLEANING Soot and scale may build up inside the chamber This must be removed at regular intervals as...

Page 10: ...s will identify the fault and a spare part can be despatched All furnaces have their own records so please quote the serial number model type and voltage that can be found on the rating plate on the b...

Page 11: ...off Furnace overheats The heat on light stays on all the time Heating element Thermocouple SSR switch Door switch Safety contactor Controller Thermal fuse SSR switch Over temp alarm controller Safety...

Page 12: ...8 0 STANDARD WIRING DIAGRAM...

Page 13: ...8 1 OVERTEMPERATURE WIRING DIAGRAM...

Page 14: ...W CHAMBER SIZE H W D CAPACITY LITRES NET WEIGHT Kg MRF1 1300 2 5 120 180 180 3 9 44 MRF2 1300 3 5 120 180 355 7 7 57 MRF3 1300 7 205 230 460 21 7 120 MRF4 1300 8 205 305 460 28 8 130 MRF5 1300 9 305 3...

Page 15: ...for laboratory and industrial use Please visit our web site at www vecstar com or contact us at the address below for more information UK Contact Vecstar Limited Units 11 12 Foxwood Road Dunston Tradi...