Trouble shooting 

 
Valve does not close/open:   

Power available? 

   

 

Compressed air available? 

   

 

Solenoid defective? 

   

 

Check voltage! 

   

 

Check air pressure! 

 
Leak at gate: 

 

Clean valve seat and gate! 

   

 

Replace gate if damaged! 

   

 

Correct air pressure? 

 
Leak at body: 

 

Flanges leaktight? 

   

 

Replace bonnet seal! 

 
 
 

4.1 

Global leak at bonnet seal after heating the valve! 

 
Due to a possible settlement of screw and VATSEAL it could happen that after a heat up and cool down  procedure the 
bonnet seal becomes slight leaky. In this case please check the torque (see below) of bonnet screw (see page 13 item 15) 
and tighten the screws if necessary with a torque wrench. 
 
Remark: 
Do tighten screws crosswise! 
Necessary torque of screws is for DN16 -DN50; 10 Nm / 7.4 lbf 

.

 ft. 

 
If you need any further information, please contact one of our service centers. You can find the addresses on our website: 
http://www.vat.ch

 

 

 
 
 

Maintenance & repairs 

 
 
 
Under  clean  operating  conditions,  the  valve  does  not  require  any  maintenance  during  the  specified  cycle  life. 
Contamination from the process may influence the function and requires more frequent maintenance. 
 
Before  carrying  out  any  maintenance  or  repairs,  please  contact  VAT.  It  has  to  be  individually  decided  whether  the 
maintenance/repair  can  be  performed  by  the  customer  or  has  to  be  carried  out  by  VAT.  The  fabrication  number  on  the 
valve 
 

made in Switzerland
Fabrication No.:

patented

. .

. . . . . – . . . . – . . . . / . . . .

A

 – . . . . . .

Fabrication number

 

 
has always to be specified.  
 
All supplies (e. g. compressed air, electrical power) must be disconnected for removal/installation of the valve from/into the 
system and for maintenance work. 
 

Summary of Contents for 01 Series

Page 1: ...s indicated on each product as per the label below or similar made in Switzerland Fabrication No patented 010 A Fabrication number Explanation of symbols Read declaration carefully before you start an...

Page 2: ...s or any other reproduction processes nor may it be manipulated with electronic systems duplicated or distributed without written permission from VAT Offenders are liable to pay damages The original V...

Page 3: ...ressed air connection 5 2 4 2 Compressed air connection Solenoid separate without solenoid 5 2 4 3 Electrical connection 6 3 Operation 7 3 1 Normal operation 7 3 2 Operation of manual actuator 7 3 3 O...

Page 4: ...sed Manual actuator 250 C Pneumatic actuator 200 C Position indicator 80 C Solenoid 50 C Position indicator Contact rating 5 A 250 V AC 3 A 50 V DC Solenoid see tag on solenoid The temperatures specif...

Page 5: ...essed air connection Connect compressed air only if valve has been installed into the vacuum system moving parts cannot be touched With solenoid Connect compressed air to IN internal thread R 1 8 1 8...

Page 6: ...lenoid Sockets for position indicator and solenoid are supplied with the valve Wire solenoid and position indicator according to the following diagrams Standard solenoid Solenoid for impulse actuation...

Page 7: ...manual actuator by one hand A higher torque may damage the gate seal 3 3 Operation under increased temperature See Technical data The temperatures specified in the technical data are only admissible...

Page 8: ...41 Pneumatic actuator single acting with closing spring NC Emergency operation push button Press push button valve closes Release push button valve opens Emergency operation slotted screw Turn in clo...

Page 9: ...hten the screws if necessary with a torque wrench Remark Do tighten screws crosswise Necessary torque of screws is for DN16 DN50 10 Nm 7 4 lbf ft If you need any further information please contact one...

Page 10: ...onnect compressed air 4 Disconnect electrical power 5 Disconnect compressed air line and plug 4 from solenoid and plug 2 from position indicator 6 Remove screws 15 from bonnet 7 Withdraw actuator gate...

Page 11: ...in brackets refer to the drawing on page 13 1 Press out pin 9 by means of tool 8 2 Take new gate 6 and press in new pin supplied with new gate by using the same tool ATTENTION Inspect position of pin...

Page 12: ...uble acting pneumatic actuator Move gate to the intermediate position by compressed air Gate valve with single acting pneumatic cylinder and closing spring Move gate to the intermediate position by co...

Page 13: ...rating and Maintenance Instructions Series 01 DN 16 50 5 8 2 VAT Vakuumventile AG CH 9469 Haag Switzerland Tel 41 81 771 61 61 Fax 41 81 771 48 30 CH vatvalve com www vatvalve com 271691EB 2009 03 09...

Page 14: ...seal 7 91984 R1 6 9 Gate with pin 91983 R1 7 Bonnet seal 65209 01 1 Position indicator Ordering No 010 E22 42 valve with closing spring NC 69250 R1 Ordering No 010 E21 41 valve with opening spring NO...

Page 15: ...xpense whether consequential special incidental direct or otherwise caused by arising out of or connected with the manufacture delivery including any delay in or failure to deliver packaging storage o...

Reviews: