Chapter 6. CRAMPS/Multipulse Module Tests
50
UNITY
INOVA Solids Hardware Installation
01-999044-00 C0501
7.
Enter
ga
, record the spectra, and measure the residual splitting of the dioxane triplet.
shows a spectrum.
8.
Use
h2cal
to calculate
γ
B
2
. No specification exists, but a value of about 25 kHz is
needed for the 100 W test to be successful.
9.
Set HI BAND
to
HI POWER. Be sure that
dpwr=63
is calibrated for about 100 W
at the HI POWER output, or use the value of
dpwr
that gives about 100 W.
10. Enter
ga
, record the spectra, and measure the residual splitting of the dioxane triplet
with
dof=–7e4,7e4
.
CAUTION:
Avoid damaging the probe, follow all pulse duration, pulse power, and
duty cycle cautions and warnings when given for a particular probe.
Generally, pulse durations can be increased with a proportional
decrease in pulse power.
11. Use
h2cal
to calculate
γ
B
2
. Set
dpwr
to achieve a
γ
B
2
that is as close as possible
to the minimum values listed in the following table. Do not exceed this power level
for any other experiment. For the XPOLAR pulse sequence, the maximum
level1
and
level2
correspond to this
dpwr
value.
Magnet
Varian RT CP/MAS
Doty CP/MAS
Varian VT CP/MAS
200 MHz
75 kHz
—
—
Figure 21. Decoupler Field Strength
γ
B
2
ACQUISITION
SAMPLE
PROCESSING
FLAGS
sfrq
100.57
date
Apr 1 lb
10.00
il
n
tn
C13
file
exp
sb
not used
toss
n
at
0.100
DECOUPLING
gf
not used
pdp
n
np
6016
dn
H1
awc
not used
dp
y
sw
3000.0
dof
arrayed
lsfid
not used
SPECIAL
xpol
n
dm
nny
phfid
not used
srate
415
tpwr
55
dmm
c
wtfile
temp
not used
tpwrf
4095
dpwr
63
proc
ft
pw
8.0
level1
45
fn
32768
p1
0
level1f
200
FLAGS
d1
30.000
level2
50
werr
ai no ph
tof
0
level2f
4095
wexp
nt
1
dpwrf
4095
wbs
ct
1
wnt
Gamma-H2 calculation
See J. Magnetic Resonance 7:442 (1972)
jo = 284.00 Hz,
dof[1] = –70000.00 Hz,
dof[2] = 70000.00 Hz
jr[1] = 165.00 Hz,
jr[2] = 167.00 Hz
Gamma-H2
=
69404.97 Hz
90 degree pulse =
3.6 usec
coalescence frequency =
–457.51 Hz
Summary of Contents for UNITY INOVA NMR
Page 13: ...Introduction 12 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 17: ...Chapter 1 Getting Started 16 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 21: ...Chapter 2 Wideband NMR Module 20 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 65: ...Index 64 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...