01-999044-00 C0501
UNITY
INOVA Solids Hardware Installation
43
Chapter 6.
CRAMPS/Multipulse Module Tests
The CRAMPS/Multipulse Solids Module enables Varian
UNITY
INOVA spectrometers to run
CRAMPS and multipulse experiments. The CRAMPS/multipulse experiment is normally
used to observe
1
H. In CP/MAS, the
1
H amplifier is used as a heteronuclear decoupler,
while in CRAMPS the
1
H amplifier is used as an high-power observe channel. This chapter
contains system tests to check the performance of the CRAMPS/Multipulse module. The
following is a list of the tests that are described:
RF transmitter performance tests
•
1 kW amplifier tuning
•
Liquids observe transmitter rf signal path
•
Solids observe transmitter rf signal path
•
Third cabinet status light and safety switch performance
•
AMT amplifier calibration
•
1 kW amplifier calibration gating
CRAMPS/Multipulse NMR tests
•
Decoupler calibration
•
FLIPFLIP
•
FLIPFLOP
•
MREV8
•
1
H wideline
CRAMPS probe test (Varian-supplied probe only)
•
MREV8 narrowing at magic angle
•
BR24 narrowing at the magic angle
CAUTION:
Read the following cautions:
• Make sure that all amplifiers are terminated in appropriate 50-ohm
loads at all times.
• The high power available from the amplifier in the third cabinet is
intended only for solid state. Applying this power to a liquids probe
will destroy the probe. Before turning on the amplifier in the third
cabinet, check that the appropriate solids probe has been put in place.
• Disable all high-power solids amplifiers before installing any probe.
• Confirm that the rf signal being measured can never exceed the
maximum input signal of the measuring device.
• Do not block any of the ventilation holes in the third cabinet. Even on
standby, the amplifiers radiate considerable heat.
Summary of Contents for UNITY INOVA NMR
Page 13: ...Introduction 12 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 17: ...Chapter 1 Getting Started 16 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 21: ...Chapter 2 Wideband NMR Module 20 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 65: ...Index 64 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...