Chapter 5. Wideline Module Tests
34
UNITY
INOVA Solids Hardware Installation
01-999044-00 C0501
absolute value mode, with
vs=1000
, normalized,
nm
. Confirm that the quadrature
image is less than 1% by height as compared to the main resonance.
8.
Set
tof
to 1 kHz off-resonance.
9.
Measure
2
H sensitivity. Use a spectral noise region of 20 kHz, centered
approximately 10 kHz off-resonance. Refer to
for specifications.
Figure 8. Sample Spectrum, S/N = 249.6, 300-MHz System
1000 kHz
800
700
600
500
400
300
200
100
0
S/N 249.6
ACQUISITION
SAMPLE
PROCESSING
FLAGS
sfrq
61.395
date
Apr 1 lb
50.00
il
n
tn
H2
solvent
none
sb
not used
in
n
at
0.008
file
/vnmr/~
gf
not used
dp
y
np
16384
stdpar/H2
awc
not used
hs
nn
sw
1e+06
DECOUPLING
lsfid
not used
fb
1e6
dn
H1
phfid
90.0
bs
not used
dof
0
wtfile
ss
0
dm
n
proc
ft
SPECIAL
tpwr
57
dmm
c
fn
65536
temp
not used
pw
2.2
dmf
4000
math
f
p1
0
homo
n
d1
4.000
werr
d2
0
wexp
tof
0
wbs
nt
1
wnt
ct
0
Table 4. 90° Pulse and Signal-to-Noise Specifications
System
Nucleus
PW90
Frequency
S/N
400 MHz
2H
≤
3.0 µs
61.4
300:1
23Na
≤
3.0 µs
105.8
630:1
300 MHz
2H
≤
2.5 µs
46.1
200:1
23Na
≤
2.5 µs
79.4
475:1
200 MHz
2H
≤
2.2 µs
30.7
110:1
23Na
≤
2.2 µs
52.9
315:1
Summary of Contents for UNITY INOVA NMR
Page 13: ...Introduction 12 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 17: ...Chapter 1 Getting Started 16 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 21: ...Chapter 2 Wideband NMR Module 20 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 65: ...Index 64 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...