5.2 Wideline NMR Tests
01-999044-00 C0501
UNITY
INOVA Solids Hardware Installation
35
Sodium 90° Pulse and Sensitivity Measurement
Use the standard liquid sample(00-969104-03), the same as for
2
H.
1.
Enter
rtp('/vnmr/tests/Na23wl')
to recall parameters.
1.
Set up the spectrometer for data acquisition as follows:
seqfil='s2pul'
nt=1 sw=1e5 np=max dm='n' dp='y' lb=50 gain=0 rof1=30
alfa=0 fn=65536
.
2.
Tune the probe to the sodium resonance frequency (refer to the NMR Probes
Installation Manual).
3.
Determine the 90° pulse for
23
Na. Enter this value for
pw90
.
4.
Set
tof
to 1 kHz off-resonance.
5.
Measure
23
Na sensitivity. Use a spectral noise region of 20 kHz, centered
approximately 10 kHz off-resonance. Refer to
Figure 9. Sample Sodium 90° Pulse and Sensitivity Spectrum, 300-MHz System
1000 kHz
500
0
ACQUISITION
SAMPLE
PROCESSING
FLAGS
sfrq
105.800
date
Apr 1 lb
50.00
il
n
tn
Na23
solvent
none
sb
not used
in
n
at
0.004
file
/vnmr/~
gf
not used
dp
y
np
16384
tests/Na23wl
awc
not used
hs
nn
sw
2e+06
DECOUPLING
lsfid
not used
fb
1e+06
dn
H1
phfid
90.0
bs
not used
dof
0
wtfile
ss
0
dm
n
proc
ft
SPECIAL
tpwr
57
dmm
c
fn
65536
temp
not used
pw
2.2
dmf
4000
math
f
p1
0
homo
n
d1
4.000
werr
d2
0
wexp
tof
0
wbs
nt
1
wnt
ct
0
Summary of Contents for UNITY INOVA NMR
Page 13: ...Introduction 12 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 17: ...Chapter 1 Getting Started 16 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 21: ...Chapter 2 Wideband NMR Module 20 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...
Page 65: ...Index 64 UNITYINOVA Solids Hardware Installation 01 999044 00 C0501 ...