TECHNICAL INFORMATION
64
TURBO-V 300 ICE CONTROLLER - 87-900-902-01 (G)
•
Choose the desired selection by pressing the CYCLE
push-button.
F R O N T / R E M O T E / S E R
S E L E C T I O N : F R O N T
•
Press PUMP CURRENT push-button and the display
shows
S E R I A L
A D D R E S S
X X
This message will only be displayed if the RS 485 module
is installed.
Where:
XX = 00
÷
31 is the unit address
•
Press the PUMP CURRENT push-button to enter the
value and ,if the serial option is installed, the display
shows:
R S X X X
B A U D
R A T E
X X X X
where:BAUD RATE
XXXX
= means 600, 1200, 2400,
4800, 9600 baud rate for the host computer or printer
communication. The controller is factory-set for 9600 baud
rate operation.
The value which corresponds to the board installed will be
assigned to RSXXX.
•
If the RS 232 is installed, enter the value by pressing
the PUMP CURRENT push-button and, if the serial
option is installed, the display shows:
R S X X X
H O S T / P R I N T
S E L E C T I O N :
X X X X X
This message will only be displayed if the RS232 interface
is present.
where:SELECTION:
XXXXX
= means HOST or PRINT.
The controller is factory-set to HOST.
Summary of Contents for Turbo-V 300 ICE