TECHNICAL INFORMATION
36
87-900-915-01(E)
ing temperature sensor, pin E is connected to the
pump ground and pins J, K and L are spare.
PUMP USED IN PRESENCE OF MAGNETIC
FIELDS
Magnetic fields induce eddy currents in the rotor of
a turbomolecular pump that tend to oppose to its
rotation. The result is increased electrical power
consumption by the motor, most of which is dissi-
pated in the rotor.
Since the rotor is not in contact with the stator the
above power can leave the rotor mainly by radia-
tion and hence the rotor may be overheated while
static parts of the pump remain cool. This effect is
strongly dependant from the intensity, time func-
tion and distribution of the magnetic field. In gen-
eral, therefore, an increase in pump current can be
expected.
If this increase is lower than 50% of the current
value drawn by the motor in high vacuum opera-
tion, no particular problem should be expected.
However if the effect is grater, than the case
should be carefully reviewed by Varian's specialist.
As a matter of fact, in case of high magnetic fields,
also important forces might be generated and ap-
plied to the rotor.
If one temperature sensor is disconnected, the
pump will not start. To prevent damage to the
pump when the temperature exceeds 60° C, the
sensor automatically cuts out the power supply.
MOVING THE GAS PURGE VALVE AND
FOREVACUUM PUMP CONNECTIONS
All the pneumatic connections to the Turbo-V2000
ICE pump can be moved to its bottom (see the fol-
lowing figure).
ACCESSORIES
DESCRIPTION PART
NUMBER
Inlet screen, ISO 250
969-9350
Heating jacket, 120 Vac
969-9821
Heating jacket, 220 Vac
969-9820
Water cooling kit for American fractional
tubes
969-9338
Water cooling kit for Metric tubes
969-9348
Pump to controller cable
969-9954
Vent flange NW 10 KF (M8)
969-9108
Vent valve with fixed delay time
969-9843
Vent device with adjustable delay time
969-9831
Heavy duty vent valve
969-9842
Solid state flowmeter 1/4" Swagelock
969-9114
Solid state flowmeter KF16
969-9115
Monitor autodiagnostic unit
969-9252
Purge vent device
969-9116
Triscroll Dry Vacuum Pump PTS 600
single phase, universal motor
PTS06001UNIV
Triscroll Dry Vacuum Pump PTS 600 3
phase, universal motor
PTS06003UNIV
Dual stage rotary vane pump DS 602
with single phase motor
949-9335
Dual stage rotary vane pump DS 602
with 3 phase motor
949-9336
Dual stage rotary vane pump DS 1002
with 3 phase motor
949-9340
To move the gas purge valve to the bottom of the
pump make the following steps:
1. Unscrew the gas purge valve from the side of
the pump
2. Unscrew the purge port cover on the bottom of
the pump and screw it on the side
3. Screw the gas purge valve (with a torque of 2.5
Nm) in the bottom of the pump.
To move the forevacuum pump flange to the bot-
tom of the pump make the following steps:
1. Unscrew the three screws that fix the forevac-
uum blank flange to the bottom of the pump
2. Unscrew the three screws that fix the forevac-
uum flange to the side of the pump
3. Fix the forevacuum flange to the bottom of the
pump using the relevant three screws
4. Fix the forevacuum blank flange to the side of
the pump using the relevant three screws.
For a complete overview of Varian's extensive