ProStar 325 UV-Vis Detector Control Manual
-
17 -
03-914947-44:2
In the
Peak Relay
column you can also activate the Pulse Relay
function The Pulse relay activates or deactivates the Peaksense
relay at the specified time. This parameter cannot be set at initial
conditions (time 0.00).
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
NOTE:
For more details about the Peak Sensor functions (Timeslice or Pulse), refer to
the appendix (
Peak Sensor functions
section) at the end of this manual.