ProStar 325 UV-Vis Detector Control Manual
-
22 -
03-914947-44:2
The following screen is displayed:
In the
Acquisition Parameters
tab, required fields like the
Run Name
, the
Run ID
and the
Acquisition length
must be
filled in before starting an analysis. Furthermore, if the
system is equipped with an autosampler controlled by
Galaxie Chromatography Data System, the
Injection
Volume
, the
Vial
position and the
Rack
number (if required)
must be entered.
Other parameters associated with each acquisition channel
can be entered in the
Channel X
tab.
3. Click on the
Start
button to start the acquisition if the
system is a single-injector system or click on the
Next
button if the system is a multiple injector system. In a
multiple injector system, the same parameters entered for
the first injector must be entered for the other injectors
before starting the acquisition.