INSTRUCTION MANUAL
8
87-900-978-01(D)
CAUTION
Check that your electrical mains voltage corresponds to that
indicated on the motor’s plate.
Connect the pump to the power supply.
USE
Before being put into service, in order to reach maximum vac-
uum, the pump must be left running for about an hour with the
gas ballast valve open. This will eliminate any humidity from the
oil.
There are no special procedures for switching the pump on; it
needs only to be connected to the electric power by means of
the bipolar switch.
,
WARNING!
The pump is designed for operation with neutral or non-
corrosive fluids. It is absolutely forbidden to use potentially ex-
plosive or flammable substances.
MAINTENANCE
Personnel responsible for pump operation and maintenance
must be well-trained and must be aware of the accident preven-
tion rules.
,
WARNING!
Death may result from contact with high voltages. Always take
extreme care and observe the accident prevention regulations
in force.
,
WARNING!
When machine is powered take care on account of moving
parts and high voltages.
,
WARNING!
If you have to perform maintenance on the pump after a con-
siderable time in operation, leave it to cool as temperature of
the outer surface may be in excess of 60 °C.
,
WARNING!
Always disconnect the power supply to the pump before start-
ing maintenance work. Place a special warning signs over the
power supply breaker switch: MACHINE UNDERGOING
MAINTENANCE - DO NOT POWER ON. When finished, re-
move the safety warning.
,
WARNING!
Do not change the oil immediately after stopping the machine
as the oil temperature may still be high.
NOTE
Before returning the pump to the constructor for repairs the
"Health and Safety" sheet attached to this instruction manual
must be filled-in and sent to the local sales office. A copy of the
sheet must be inserted in the pump package before shipping.
If a pump is to be scrapped, it must be disposed of in accor-
dance with the specific national standards.
DISPOSAL
Meaning of the "WEEE" logo found in labels
The following symbol is applied in accordance with the EC
WEEE (Waste Electrical and Electronic Equipment) Directive.
This symbol (
valid only in countries of the European Com-
munity
) indicates that the product it applies to must NOT be
disposed of together with ordinary domestic or industrial waste
but must be sent to a differentiated waste collection system.
The end user is therefore invited to contact the supplier of the
device, whether the Parent Company or a retailer, to initiate the
collection and disposal process after checking the contractual
terms and conditions of sale.