TECHNICAL INFORMATION
TECHNICAL DATA
The following table lists the main technical data of the HS452 and HS652 pumps.
TECHNICAL DATA
Hz
UNITS
HS452
HS652
FREE AIR DISPLACEMENT
66
l/min (m
3
/h)
450 (27)
658 (39.5)
46
l/min
(m
3
/h)
327 (19.6)
478 (28.7)
PUMPING SPEED (at 5 mbar inlet pressure)
66
m
3
/h 21 29.5
46
m
3
/h 14.5 20.5
ULTIMATE PARTIAL PRESSURE *
mbar
10
-4
10
-4
ULTIMATE TOTAL PRESSURE *
mbar
2
⋅
10
-3
2
⋅
10
-3
ULTIMATE TOTAL PRESSURE WITH GAS BALLAST *
mbar
1
⋅
10
-2
1
⋅
10
-2
WATER VAPOR TOLERANCE
mbar
30
30
WATER VAPOR CAPACITY
g/h
550
550
OIL CAPACITY min/max
l
0.6/1
0.6/1
MOTOR RATING 3ph
kW
0.5
0.5
NOMINAL ROTATION SPEED (low speed)
66
rpm
1980
1980
46
rpm
1350
1350
NOISE LEVEL with gas ballast closed
dB(A)
≤
60
≤
60
OIL TEMPERATURE
°C
75
75
(pump operating) **
°F
169
169
Installation category
II
Pollution degree
2
OPERATING TEMPERATURE RANGE
°C
12 - 40
12 - 40
WEIGHT
Kg
33
33
lb
72.7
72.7
INLET FLANGE
DN
25KF
25KF
EXHAUST FLANGE
DN
25KF
25KF
Main Dimensions:
- length
mm
515
515
- width
mm
164
164
- height
mm
271
271
Nominal Input Voltage
V
100 – 120
200 - 240
100 – 120
200 - 240
Input frequency
Hz
50 - 60
50 – 60
Max input power
VA
1000
1000
Main Fuse (T type)
A
10
10
* According to PNEUROP 6602
** At ultimate total pressure, 20 °C (68 °F) room temperature