TECHNICAL INFORMATION
70/74
87-901-010-01
Window Meanings
N. Read/
Write
Data
Type
Description Admitted
Values
000 R/W L Start/Stop
(in remote/
mode the win-
dow is a read
only)
Start = 1
Stop = 0
001 R/W L Low
Speed
Activation (in
remote/ mode
the window is
a read only)
No = 0
Yes = 1
(default = 0)
008 R/W L Remote
or Serial con-
figuration
Remote = 1
Serial = 0
(default = 1)
100 R/W L Soft
Start
(write only
in Stop
condition)
YES = 1
NO = 0
101
R/W
N
R1 Set Point
type
0 = Frequency
1 = Power
2 = Time
3 = Normal
4 =Pressure
(default = 0)
102
R/W
N
R1 Set Point
valve
(expressed in
Hz, W or s)
(default = 867)
103
R/W
N
Set Point de-
lay: time be-
tween the
pump start and
the set point
check (sec-
onds)
0 to 999999
(default = 0)
104
R/W
L
Set Point sig-
nal activation
type: the signal
can be "high
level active" or
"low level ac-
tive"
0 = high level ac-
tive
1 = low level ac-
tive
(default = 0)
105 R/W N Set
point
hysteresis
(in % of value)
0 to 100
(default = 2)
106 R/W L Water
cooling
(write only in
stop)
0 = NO
1 = YES
107 R/W L Active
Stop
(write only in
stop)
0 = NO
1 = YES
(default = 0)
108
R/W
N
Baud rate
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
109 W L Pump
life/
cycle time/
cycle number
reset
To reset write ‘1’
110
R/W
L
Interlock type
(default = 1)
Impulse = 0
Continuous = 1
N. Read/
Write
Data
Type
Description Admitted
Values
111 R/W L Analog
out-
put type:
output volt-
age signal
proportional
to frequency
or power
0 = frequency
1 = power
(default = 1)
2 = Pump
temperature
3 = Pressure
117 R/W N Low
Speed
frequency
(Hz)
1100 to “Maxi-
mum rotational
frequency”
(win 120)
(default = 1100)
120 R/W N Rotational
frequency
setting (Hz)
1100 to 1350
(default = 1350)
122 R/W L Set
vent
valve on/off
0 = Closed
1 = Open
(default = 1)
125
R/W
L
Set the vent
valve
operation
Automatic = 0
(see note 1.)
On command = 1
(see note 2.)
126 R/W N Vent
valve
opening delay
(expressed in
0.2 sec)
0 to 65535
(corresponding to
0 to 13107 sec)
(default = 15)
147 R/W N Vent
open
time
See “vent
connector”
paragraph
0 = infinite
1 bit = 0.2 sec
1 to 65535
155 R N Power
limit
applied
Read the
maximum
allowable
power
watt
157 R/W N Gas
load
type
Select the
gas load to
the pump
Ø = N
2
1 = Ar
161 R/W N Pressure
Reading Con-
nection Factor
Table A1 Ap-
pendix A Eye-
Sys Manual
0 to 10
0 = 0
10 = 1 = N
2
162
R/W
A
R1 Set Point
Pressure
Valve
Valid if Win.
101 = 4
Format X.X
E. XX
Where X = 0
to 9
s = + or -
163 R/W N Pressure
unit
of measure
0 = mBar
1 = Pa
2 = Torr