![Varian 969-8217 Instruction Manual Download Page 65](http://html.mh-extra.com/html/varian/969-8217/969-8217_instruction-manual_854731065.webp)
TECHNICAL INFORMATION
60/74
87-901-010-01
INLET SCREEN INSTALLATION
The inlet screens mod. 969-9300, 969-9309,
969-9302 and 969-9303 prevent the blades of the
pump from being damaged by debris greater than
0.7 mm diameter.
The inlet screen, however, does reduce the pump-
ing speed by about 10%.
The inlet screen is fitted in the upper part of the
pump, as shown in the figure.
The screen can be mounted on each pump.
The screen can be removed as shown in the fol-
lowing figure.
The following figure shows the overall flange di-
mensions with the protection screen fitted on
pump with ISO flange and pump with CFF flange
(dimensions are in inches [mm]).
0.2
4 [6.0
0]
0.
35 [
9.0
0]
2.76 [70.00]
2.09 [53.00]
ISO 63 / CFF 2.75 / CFF 4.5
KF 40
0.
1 [2.5
0]
DIMENSIONS: inches [mm]
CFF 8"
ISO 160
ISO 100
CFF 6"
0.67
[1
7.00]
0.47
[12.00
]