Table 6 Spatial Setup Recommendations for 1200L LC/MS and GC/MS
Recommendations
Purpose
Recommended Distance
Allow adequate space to the side of
the LC.
Solvent bottle placement for
HPLC pumps. See Figure 3.
≥
24 in. (61 cm)
Allow adequate space to the side of
the GC/MS.
Permit access to the transfer
line. See Figure 1.
~ 24 in. (61 cm)
Allow adequate space behind the
system.
Provide clear space for air
circulation, gas lines, and
electrical connections.
6 to 12 in. (15 to 30 cm)
Allow vertical clearance above the
instrument.
Dissipate heat and allow for
routine maintenance.
≥
30 in. (76 cm)
Allow additional bench space beside
the Workstation keyboard.
Enable mouse operation.
8 in. x 8 in. (20 cm x 20 cm)
Appendix 2. Power Consumption
Table 7 Specifications for maximum power draws for UPS systems
(uninterruptible power supply)
MS Configuration
Maximum Power Draw
1200L LCMS without pumps
1200 VA
1200L GCMS with DS102 pump
1200 VA
1200 with DS102 pump
1200 VA
1200L LCMS Pumps (2 DS602 pumps)
3072 VA, 5280 VA surge < 30 sec,
10900 VA surge < 1 sec
03-914961-00:6