Vapotherm 2000I Operating Instruction Manual: Appendix B: Sample Disinfection Log Sheet
Page 27
SAMPLE
Vapotherm 2000i
Operating Instruction Manual
Appendix B
– Sample Disinfection Log
Vapotherm 2000i
Hospital ID
Disinfection
Disinfectant
Operator
Machine Serial No.
(If Applicable)
Date
Solution &
Hold Time
Vapotherm 2000I Operating Instruction Manual: Appendix A Disinfection Solutions
Page 26
Vapotherm 2000i
Operating Instruction Manual
Appendix A
– Disinfection Solutions
(cont.)
Dilution Instructions for preparing a 1% solution of Minncare
®
and Control III
™
Supplies Needed:
• A beaker, graduated cylinder, or container suitable for measuring and mixing 200ml of
disinfectant solution
WARNING:
Components or instruments that are not part of the DK-301 Kit and are used to mix
or pour disinfectant solutions should be either single use components or glassware that has been
sterilized, high level disinfected, or pasteurized before use in accordance with your institution’s
specific disinfection and sterilization procedures.
• 10ml syringe, or pipette, graduated in 1ml increments
• Concentrate Disinfectant
• Minncare
®
Cold Sterilant EPA Reg No. 52252-4 (22.00% Hydrogen Peroxide, 4.50%
Peroxyacetic Acid, 73.50% Inert Ingredients)
•
Control III
®
Disinfectant P/N 10006 (10% n-alkyl dimethyl benzyl ammonium chloride,
10% n-aykyl ethyl benzyl ammonium chloride, 80% inert ingredients)
• 200
ml of sterile water
WARNINGS:
Always follow manufactures recommendations for handling disinfectant solutions.
Always wear Personal Protective Equipment when handing disinfectant solution.
Procedure:
1. Don appropriate Personal Protective Equipment: Safety goggles, gloves, splash apron.
2. Using an appropriate syringe or pipette, draw up 2
ml of undiluted Minncare
®
(EPA Reg No. 52252-4) or Control III
®
(P/N 10006) and expel it in a device appropriate for
measuring 200
ml of solution.
3. Make up 200
ml of solution by adding sterile water to the measuring device.
4. Close the slide clamp on Disinfection Bag A and pour disinfectant solution into the bag.
5. Close the cap of Bag A firmly, to avoid spills and continue with Step 6 in Section 8.4 of the
disinfection procedure (Page 18).
6. Attach to IV Pole and Vapotherm 2000i according to Disinfection procedure and continue to
follow Disinfection Protocol.