9. HYGIENE

9.1. General  

Caution!
To avoid a health risk – e.g. in-

fection from a contaminated 

Filling Chamber or a contaminated 

Valve Balloon – the following hygiene 

instructions must be followed.  
In a physician‘s practice or a in a clinic, 
the hygiene requirements of the RKI 

(Robert Koch Institute), the BfArM 
(German Federal Institute for Drugs 
and Medical Devices) or comparable 

institutions in your country, have to be 

observed. 

The 

VOLCANO MEDIC

 Vaporization 

System is intended for multiple use.  
a) Filling Chamber

  The Filling Chamber must be 

cleaned if there is visible residue 
or when a new Valve Balloon is 
used.

b) Valve Balloon 

  The Valve Balloon may be used by 

one person only.

Make sure that the equipment is al-
lowed to dry thoroughly after every 
cleaning. Condensation or residual 
moisture may encourage bacterial 

growth. Please note Chapter 9.2. “Ma-

terial Resistance” on this page. 
Inspect the component parts of the 
Filling Chamber regularly and replace 

defective (broken, misshapen, dis-
colored) parts. Replace the Valve Bal-

loon after 70 applications or at least 

14 days after the first use.  

The Lip Piece of the 

VOLCANO 

MEDIC

 Vaporization System is 

equipped with a check valve. 

This check valve reduces the danger of a 

potential contamination inside the Valve 
Balloon by preventing breathing into the 

Valve Balloon. 

This ensures that the Valve Balloon can 

be re-used several times. 

9.2.  Material Resistance of the 

VOLCANO MEDIC

 Vaporiza-

tion System 

Please ensure that the Mouthpiece, 
the Filling Chamber, the Lip Piece and 
the Hot Air Generator do not come in 
contact with the following chemicals:

  organic, mineral and oxidating 

acids (min. permissible pH 5.5) 

  strong bases (max. permissible 

pH 10). We recommend using 
neutral/enzymatic or slightly 
alkaline cleaners.  

  organic solvents (e.g. aldehyde, 

ether, ketone, benzene)  

  oxidation agents (e.g. hydrogen 

peroxide) 

  halogens (chlorine, iodine, bro-

mine)  

  aromatic/halogenated hydrocar-

bons

9. HYGIENE

  phenol derivatives, amin com-

pounds (in high concentrations)

  oils and greases

Clean the parts of the 

VOLCANO 

MEDIC

 Vaporization System only 

with soft brushes, never with metal 
brushes or steel wool. 

9.3.   Disposable 

MEDIC VALVE

  

Lip Piece for Hygienic Use

The Lip Piece will be delivered brand 

new and in a low-germ, but not fully-
sterile state. 

The Lip Piece is equipped 

with a check valve. This 
check valve reduces the dan-

ger of a potential contamination inside 

the Valve Balloon by preventing 
breathing into the Valve Balloon. 

This ensures that the Valve Balloon 

can be re-used several times before 

it has to be disposed (see Chapter 
9.4. “Reusability of the Valve Balloon”, 
page 41). 

Valve Balloon with Mouthpiece and Lip 
Piece

The Lip Piece is put on the Mouth-

piece, slid until stop and is then ready 
for operation. 

The Valve Balloon with Mouthpiece 

and the Lip Piece may be used by only 
one person. 

A Lip Piece may be used for max. 4 

hours after initial operation if it is un-

damaged and clean. Afterwards, it 
must be disposed of in the household 
waste and a new Lip Piece must be 
used due to hygienic reasons. 
Every further use beyond this or the 
use of damaged and/or contaminated 
products respectively is the user‘s re-
sponsibility. 
We assume no liability if this recom-
mendation is ignored. 

Please note that damaged or 

brittle Lip Pieces are not to be 

used.

Summary of Contents for VOLCANO MEDIC

Page 1: ...INSTRUCTIONS FOR USE Read carefully before using the device and save these instructions...

Page 2: ...e and Transport Conditions 16 5 The VOLCANO MEDIC Hot Air Generator 17 5 1 Unpacking 17 5 2 Installation 17 5 3 Operation 17 5 4 Commissioning 18 5 5 Heating Up 18 5 6 Temperature Setting 19 6 The Use...

Page 3: ...Filling Chamber Reusability of the Filling Chamber Reusability of the Valve Balloon Storage 9 5 Disassembly and Assembly 42 Filling Chamber Filling Chamber Cap Mouthpiece Disassembling the Filling Ch...

Page 4: ...Pad 3 pcs Filling Chamber Clips not shown 1 pc Cap Ring not shown Please check whether all components have been included If not please inform our Service Center 1 PRODUCT OVERVIEW SCOPE OF DELIVERY S...

Page 5: ...during installation use and maintenance of the VOLCANO MEDIC Vaporiza tion System Please keep this brochure in a safe place for future reference You may download the current version of the Instructio...

Page 6: ...given on the type label on the bottom side of the Hot Air Generator cor responds to the mains supply at the place of installation If there are any problems during operation immediately pull out the po...

Page 7: ...or should be cleaned only with a dry cloth or a damp cloth if absolutely necessary Never im merse the Hot Air Generator in water or in other liquids or clean it with a direct water or steam jet The Ho...

Page 8: ...r responding laws Use only medicinal hemp flowers cannabis flos prescribed by your physician and purchased from a phar macy Otherwise you may be at risk of breaking the law In case of doubt ask your p...

Page 9: ...and work properly If you notice any defects inform the seller or the delivery agent immediately The packing material plastic bags styrofoam material boxes etc must be kept out of the reach of children...

Page 10: ...constantly maintained at the set value The luminous intensity of the yellow control lamp corresponds to the intensity of the heating process For technical reasons the temperature in the Filling Chamb...

Page 11: ...10 C 410 F Cannabinoid content in blood stream systemic bioavailability after inhalation 6 THE USE OF DRONABINOL Preparation to use the Liquid Pad When using dronabinol THC dis solved in alcohol the S...

Page 12: ...apter 9 5 Disassembling and As sembling the Filling Chamber page 42 Either the small Liquid Pad meas urements 15 x 5 mm or the Dosing Capsule containing a Liquid Pad measurements 13 x 6 mm are to be i...

Page 13: ...button and shut off the air pump and remove the Valve Balloon together with the Filling Chamber To do so hold the Filling Chamber Cap at the textured grip 6 THE USE OF DRONABINOL 6 4 Inhalation from t...

Page 14: ...uate considerably and they often contain herbicides fertilizers bacteria etc 7 1 The Efficiency of Cannabi noids vaporized from Hemp Flowers Cannabis Flos By following the procedure listed in the Inst...

Page 15: ...ge is too high reduce the vaporization temperature Drug Filling Chamber Content Bedrocan 100 mg Bediol 100 mg THC approx 19 mg THC approx 6 mg CBD approx 7 5mg approx 8 mg approx 3 5 mg approx 4 5 mg...

Page 16: ...ve Balloon tight Fill the Filling Chamber Reducer with ground hemp flowers directly or with a filled Dosing Capsule Take care that the Screens are not obstructed with plant material To guarantee opti...

Page 17: ...ber to overheat and may cause burns When the Valve Balloon is filled switch off the air pump and remove the Valve Balloon together with the Filling Chamber To do so hold the Filling Chamber Cap at the...

Page 18: ...the Dosing Capsule by using the Maga zine Cap funnel Dosing Capsules with Magazine With the Dosing Capsules which are available as accessories you can fill ground hemp flowers or dronabinol using the...

Page 19: ...d Liquid Pad can be filled with up to 5 drops of dronabinol and store them in the Magazine Please make sure to not overfill the Liquid Pad 8 ACCESSORIES DOSING CAPSULES AND MAGAZINE The Dosing Capsule...

Page 20: ...tor do not come in contact with the following chemicals organic mineral and oxidating acids min permissible pH 5 5 strong bases max permissible pH 10 We recommend using neutral enzymatic or slightly a...

Page 21: ...rgent 9 HYGIENE Put the component parts of the Filling Chamber in a suit able place in the dishwasher Small parts e g Filling Chamber Clips can be lost or may damage the dishwasher if improperly place...

Page 22: ...Disassembling the Filling Chamber Push the lower Screen or the Liquid Pad upwards out of the Filling Cham ber Cylinder Push the Filling Chamber Cylinder upwards out of the Filling Chamber Housing Pus...

Page 23: ...hold the de vice under running water In a physician s office or clinic the housing of the Hot Air Generator can be cleaned and disinfected with a sur face disinfectant The surface disinfectant should...

Page 24: ...OLCANO MEDIC display no activation of the heater or pump possible Please make sure that the current is on and that the fuse in the mains fuse box is switched on and is not defective If the Hot Air Gen...

Page 25: ...self Contact directly our Service Center 10 7 Periodic Safety Checks The following safety checks should be performed at least every 24 months by a qualified person who has ade quate training knowledge...

Page 26: ...ied below The customer or the user of the VOL CANO MEDIC Vaporization System should assure that it is used in such an en vironment Voltage dips short interruptions and voltage variations on power supp...

Page 27: ...onment specified below The customer or the user of the VOL CANO MEDIC Vaporization System should assure that it is used in such an en vironment Where P is the maximum output power rating of the transm...

Page 28: ...tomer or the user of the VOLCANO MEDIC Hot Air Generator can help prevent electro magnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transm...

Page 29: ...ing Capsule Set 40 pcs REF 09 36 M MEDIC VALVE Magazine with 8 Dosing Capsule for Liquids REF 09 38 M MEDIC VALVE Dosing Capsules for Liquids Set 40 pcs REF 03 35 M MEDIC VALVE Liquid Pad Set for Dosi...

Page 30: ...ring a new product to the customer Any warranty claims must be di rected to Storz Bickel GmbH In Gruben cker 5 9 78532 Tuttlingen Please send the complete defective item to us not disassembled with in...

Page 31: ...Signature Name Title OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK 61 60 13 WARRANTY LIABILITY REPAIR SERVICE 14 PROOF OF THE...

Page 32: ...STORZ BICKEL GMBH CO KG In Gruben cker 5 9 78532 Tuttlingen Germany Phone 49 7461 969707 0 Fax 49 7461 969707 7 E Mail info vapormed com by STORZ BICKEL GmbH Co KG VMAL 50 213 EN 08 2017 Subject to al...

Reviews: