5. THE 

VOLCANO MEDIC

 HOT AIR GENERATOR

5.4. Commissioning

Do not leave the device unat-

tended while it is being opera-
ted. Put the Hot Air Generator 

on a flat and stable surface. 
Check whether the mains voltage cor-
responds to the voltage rating stated 
on the device. The power cord and 

the plug must be in perfect condition. 
Only connect the Hot Air Generator to 

a properly installed  socket. The man-
ufacturer will not bear any responsibil-

ity whatsoever if this accident preven-
tion measure is not taken.

The device is safely disconnected 

from the mains only when the plug is 
removed from the socket. 
When the power plug is plugged in, 
the LED display of the device lights 
up briefly. At the same time, an inter-
nal software query is performed. The 

green dot on the display shows that 

the Hot Air Generator is connected to 
the power supply.

5.5. Heating Up  

To switch on the heater, push the but-

ton labeled “HEAT”. The set/actual 
temperature will light up, signaling 
that the heater is switched on. At the 
same time, the yellow control lamp 
will light up to indicate that the device 
is heating up. The upper, red display 

shows the actual temperature in the 
Filling Chamber with the air pump 
switched on, and the lower, green 
display shows the desired set tem-
perature. The heating process is com-
pleted when the set value is equal to 
the actual value. Heating up can take 
up to 5 minutes depending upon the 
chosen temperature. After the heating 
process is completed, the tempera-
ture is constantly maintained at the 
set value. The luminous intensity of 
the yellow control lamp corresponds 
to the intensity of the heating process.

For technical reasons the 

temperature in the Filling 
Chamber cannot be meas-

ured during vaporization. During op-
eration the temperature is measured 
at the heat exchanger. The display in-
dicates the reference values for the 
attached Filling Chamber with the air 
pump switched on.

5. THE 

VOLCANO MEDIC

 HOT AIR GENERATOR

5.6. Temperature Setting

Plus temperature 

button

Minus temperature 
button

Actual temperature 
display (red)

Set temperature  

display (green)

When you press the plus button, the 
set value is increased, and when you 
press the minus button, the set value 
is decreased. 
When you briefly tap on the plus or 
minus button, the set values are in-
creased or decreased respectively in 

1-degree increments. 

If you keep the plus or minus button 
pressed, the set values start running 
continuously in the corresponding di-
rection. 
If you simultaneously tap on the plus 
button and the minus button, the dis-
play changes from degree Celsius to 
degree Fahrenheit and vice versa.

If the actual value is higher 

than the set value, you can 
speed up the cooling down 

process by switching on the air pump. 

30 minutes after the last time 

a button is pressed, the devi-
ce automatically switches off.  

As long as the heater has a substan-

tial amount of residual warmth, the 
temperature is displayed until it drops 
below +40°C (+104°F).

Then the display turns off, and the 

green dot on the display indicates that 

the Hot Air Generator is connected to 
the mains supply.

The device is safely discon-

nected from the mains only 
when the plug is removed 

from the socket. 

Summary of Contents for VOLCANO MEDIC

Page 1: ...INSTRUCTIONS FOR USE Read carefully before using the device and save these instructions...

Page 2: ...e and Transport Conditions 16 5 The VOLCANO MEDIC Hot Air Generator 17 5 1 Unpacking 17 5 2 Installation 17 5 3 Operation 17 5 4 Commissioning 18 5 5 Heating Up 18 5 6 Temperature Setting 19 6 The Use...

Page 3: ...Filling Chamber Reusability of the Filling Chamber Reusability of the Valve Balloon Storage 9 5 Disassembly and Assembly 42 Filling Chamber Filling Chamber Cap Mouthpiece Disassembling the Filling Ch...

Page 4: ...Pad 3 pcs Filling Chamber Clips not shown 1 pc Cap Ring not shown Please check whether all components have been included If not please inform our Service Center 1 PRODUCT OVERVIEW SCOPE OF DELIVERY S...

Page 5: ...during installation use and maintenance of the VOLCANO MEDIC Vaporiza tion System Please keep this brochure in a safe place for future reference You may download the current version of the Instructio...

Page 6: ...given on the type label on the bottom side of the Hot Air Generator cor responds to the mains supply at the place of installation If there are any problems during operation immediately pull out the po...

Page 7: ...or should be cleaned only with a dry cloth or a damp cloth if absolutely necessary Never im merse the Hot Air Generator in water or in other liquids or clean it with a direct water or steam jet The Ho...

Page 8: ...r responding laws Use only medicinal hemp flowers cannabis flos prescribed by your physician and purchased from a phar macy Otherwise you may be at risk of breaking the law In case of doubt ask your p...

Page 9: ...and work properly If you notice any defects inform the seller or the delivery agent immediately The packing material plastic bags styrofoam material boxes etc must be kept out of the reach of children...

Page 10: ...constantly maintained at the set value The luminous intensity of the yellow control lamp corresponds to the intensity of the heating process For technical reasons the temperature in the Filling Chamb...

Page 11: ...10 C 410 F Cannabinoid content in blood stream systemic bioavailability after inhalation 6 THE USE OF DRONABINOL Preparation to use the Liquid Pad When using dronabinol THC dis solved in alcohol the S...

Page 12: ...apter 9 5 Disassembling and As sembling the Filling Chamber page 42 Either the small Liquid Pad meas urements 15 x 5 mm or the Dosing Capsule containing a Liquid Pad measurements 13 x 6 mm are to be i...

Page 13: ...button and shut off the air pump and remove the Valve Balloon together with the Filling Chamber To do so hold the Filling Chamber Cap at the textured grip 6 THE USE OF DRONABINOL 6 4 Inhalation from t...

Page 14: ...uate considerably and they often contain herbicides fertilizers bacteria etc 7 1 The Efficiency of Cannabi noids vaporized from Hemp Flowers Cannabis Flos By following the procedure listed in the Inst...

Page 15: ...ge is too high reduce the vaporization temperature Drug Filling Chamber Content Bedrocan 100 mg Bediol 100 mg THC approx 19 mg THC approx 6 mg CBD approx 7 5mg approx 8 mg approx 3 5 mg approx 4 5 mg...

Page 16: ...ve Balloon tight Fill the Filling Chamber Reducer with ground hemp flowers directly or with a filled Dosing Capsule Take care that the Screens are not obstructed with plant material To guarantee opti...

Page 17: ...ber to overheat and may cause burns When the Valve Balloon is filled switch off the air pump and remove the Valve Balloon together with the Filling Chamber To do so hold the Filling Chamber Cap at the...

Page 18: ...the Dosing Capsule by using the Maga zine Cap funnel Dosing Capsules with Magazine With the Dosing Capsules which are available as accessories you can fill ground hemp flowers or dronabinol using the...

Page 19: ...d Liquid Pad can be filled with up to 5 drops of dronabinol and store them in the Magazine Please make sure to not overfill the Liquid Pad 8 ACCESSORIES DOSING CAPSULES AND MAGAZINE The Dosing Capsule...

Page 20: ...tor do not come in contact with the following chemicals organic mineral and oxidating acids min permissible pH 5 5 strong bases max permissible pH 10 We recommend using neutral enzymatic or slightly a...

Page 21: ...rgent 9 HYGIENE Put the component parts of the Filling Chamber in a suit able place in the dishwasher Small parts e g Filling Chamber Clips can be lost or may damage the dishwasher if improperly place...

Page 22: ...Disassembling the Filling Chamber Push the lower Screen or the Liquid Pad upwards out of the Filling Cham ber Cylinder Push the Filling Chamber Cylinder upwards out of the Filling Chamber Housing Pus...

Page 23: ...hold the de vice under running water In a physician s office or clinic the housing of the Hot Air Generator can be cleaned and disinfected with a sur face disinfectant The surface disinfectant should...

Page 24: ...OLCANO MEDIC display no activation of the heater or pump possible Please make sure that the current is on and that the fuse in the mains fuse box is switched on and is not defective If the Hot Air Gen...

Page 25: ...self Contact directly our Service Center 10 7 Periodic Safety Checks The following safety checks should be performed at least every 24 months by a qualified person who has ade quate training knowledge...

Page 26: ...ied below The customer or the user of the VOL CANO MEDIC Vaporization System should assure that it is used in such an en vironment Voltage dips short interruptions and voltage variations on power supp...

Page 27: ...onment specified below The customer or the user of the VOL CANO MEDIC Vaporization System should assure that it is used in such an en vironment Where P is the maximum output power rating of the transm...

Page 28: ...tomer or the user of the VOLCANO MEDIC Hot Air Generator can help prevent electro magnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transm...

Page 29: ...ing Capsule Set 40 pcs REF 09 36 M MEDIC VALVE Magazine with 8 Dosing Capsule for Liquids REF 09 38 M MEDIC VALVE Dosing Capsules for Liquids Set 40 pcs REF 03 35 M MEDIC VALVE Liquid Pad Set for Dosi...

Page 30: ...ring a new product to the customer Any warranty claims must be di rected to Storz Bickel GmbH In Gruben cker 5 9 78532 Tuttlingen Please send the complete defective item to us not disassembled with in...

Page 31: ...Signature Name Title OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK OK not OK 61 60 13 WARRANTY LIABILITY REPAIR SERVICE 14 PROOF OF THE...

Page 32: ...STORZ BICKEL GMBH CO KG In Gruben cker 5 9 78532 Tuttlingen Germany Phone 49 7461 969707 0 Fax 49 7461 969707 7 E Mail info vapormed com by STORZ BICKEL GmbH Co KG VMAL 50 213 EN 08 2017 Subject to al...

Reviews: