10.3. Storage
Store the Hot Air Generator in a dry
place protected against the effects of
weather and out of the reach of chil-
dren or unqualified persons.
10.4. In Case of Problems
Immediately unplug the power cord
and ensure that no one can connect
the Hot Air Generator to the mains
supply again without this being no-
ticed.
If there are any faults in the electron-
ics of the Hot Air Generator
VOLCA-
NO MEDIC
, you can perform a reset
by pulling out the power plug and
plugging it again after waiting for mini-
mum three seconds.
The Hot Air Generator has two inter-
nal fuses (F1 and F2). As these fuses
are used only in case of a defect in
the Hot Air Generator, they can be
replaced only by our Service Center.
Put the Hot Air Generator in the origi-
nal packaging or pack it properly in
other packing material and take it to
our Service Center.
10.5. Disposal
If the Hot Air Generator is irreparably
damaged, please do not simply throw
it away.
As it is made of high quality, complete-
ly recyclable components, it should
be brought to a material recycling
centre in your town or sent to our Ser-
vice Center for disposal.
10. HOT AIR GENERATOR
Problem
Possible cause/remedy
No green light signal visible
on the VOLCANO MEDIC
display; no activation of the
heater or pump possible.
Please make sure that the current is on and that the
fuse (in the mains fuse box) is switched on and is not
defective.
If the Hot Air Generator still does not function, the inter-
nal fuses may have been triggered.
In this case, send the VOLCANO MEDIC Hot Air Generator
to our Service Center.
No aerosol is formed.
Check the set and actual temperature.
Make sure the Filling Chamber and/or Valve Balloon were
correctly attached.
Check whether dronabinol has been added to the Liquid
Pad or whether there is fresh, unused cannabis flos in the
Filling Chamber.
The Valve Balloon does not
fill up.
Check whether the pump is turned on.
Check whether the Filling Chamber and/or Valve Balloon
are correctly attached.
The device switches off
(automatic switch-off of
the VOLCANO MEDIC).
No button was pressed in the last 30 minutes – this leads
to automatic switch-off. Switch heater or pump on again.
Plant particles in the Valve
Balloon (when using can-
nabis flos).
Check whether the Screen in the Filling Chamber Cap is
clogged or is not inserted into the notch properly.
Check whether the Filling Chamber was cleaned accord-
ing to the Instructions for Use.
10. HOT AIR GENERATOR
10.6. Troubleshooting
If the device does not function properly when the power plug is connected to the
power socket, please check the Hot Air Generator using the tips in the following
table: