Vaporless ARM-4073 Installation & Operation Manual Download Page 5

Vaporless Manufacturing, Inc. 

8700 East Long Mesa Drive, Prescott Valley, Arizona   86314 

800-367-0185     928-775-5191     Fax:  928-775-5309 

Email:  [email protected]    Web Site:  

www.vaporless.com

 

2.

 

Install LDT-890 Test Unit

 

 

Select  the  dispenser  at  the  highest  point  of  the  delivery  system.  If  there  is  no  elevation 
difference, select the dispenser farthest from the turbine. 

 

Caution: Eye protection required during the calibration process. 

3.

  Carefully remove the plug from the test port on the impact valve. There should still be pressure 

in the line.

 

4.

  Install  the  18”  whip  hose  supplied  with  the  test  unit.  The  application  of  thread  sealing 

compound is recommended. 

5.

  Connect the quick disconnect coupler on the hose from the test unit to the whip hose. 

6.

  Set the test unit selector to the

 P

RESSURE 

S

TEP 

T

EST

 

position.

 

7.

 

Power On

 

 

Turn on the circuit breaker for the product under test. 

8.

 

System Purge

 

 

Authorize the dispenser. The turbine should start running. On the right-hand pressure gauge of 
the LDT-890, watch for the turbine to achieve operating pressure. 

9.

  Check all connections for leaks. Correct any fault conditions. 

10

. With  the  large  beaker  (1000  ml)  under  the  LDT-890  discharge  hose,  set  the  selector  to  the

 

D

ISPENSER 

N

OZZLE

 position. Purge the tester of air by allowing 800 to 1000 ml of fuel to flow 

into the beaker. After purging the LDT-890, set the selector to the

 P

RESSURE 

S

TEP 

T

EST

 

position. 

11.

 Purge the dispenser line by running several gallons of fuel into an approved safety container. 

12.

 

Bleed-Back Measurement

 

 

Remove the dispenser authorization. The turbine should stop running. Unless there are fault 
conditions such as a leak or significant thermal contraction, the line pressure should stabilize 
at 20 – 30 PSI for fractional horsepower turbines, or 20 – 40 PSI for high-capacity or high-
pressure turbines. 

13.

 With the large beaker under the LDT-890 discharge hose, carefully move the selector to the 

D

ISPENSER 

N

OZZLE

 position. Fuel will begin to flow into the beaker. When fuel stops flowing, 

return the selector to the

 P

RESSURE 

S

TEP 

T

EST

 

position.

 

14.

 Note the volume of fuel collected. This volume is referred to as the 

bleed-back

 measurement. 

Repeat the collection process several times to confirm the bleed-back volume. 

Depending  on  the  measured  volume,  the  use  of  the  small  (250  ml)  beaker  may  provide  more 
accurate measurements. The averaging of several consecutive samples is necessary to accurately 
determine the bleed-back volume. It is recommended that a record be made of these values and 
stored in a convenient location for future reference. 

 

 

Summary of Contents for ARM-4073

Page 1: ...Quality Petroleum Equipment Solutions for Over 30 Years ARM 4073 Automatic Repressure Module Installation Operation Manual Rev 05 08 2021 ...

Page 2: ...llowing statements outline the operation of the ARM 4073 1 With power applied the module is active and performs its timing functions The pump timer is the first to operate following the application of power Following the completion of the pump cycle the delay timer determines the elapsed time to the next pump cycle 2 There is a small switch module located in the center of the circuit board This mo...

Page 3: ...o difference whether the turbine s at your site are single phase or three phase The ARM 4073 only controls the coil on the turbine contactor just as any other authorization signal would The vast majority of installations for the ARM 4073 can be described with Figure 1 located on page 10 of this manual If you feel that your site is unique in some way please do not hesitate to consult with VMI for t...

Page 4: ...ith the lever positioned toward the terminal block power is applied to the module on position With the lever positioned away from the terminal block power is removed from the module off position Care should be exercised when operating the switch as voltage may be present on the L1 L2 terminals MEASURING SYSTEM BLEED BACK When under pressure all closed piping systems exhibit an elastic response tha...

Page 5: ...se set the selector to the DISPENSER NOZZLE position Purge the tester of air by allowing 800 to 1000 ml of fuel to flow into the beaker After purging the LDT 890 set the selector to the PRESSURE STEP TEST position 11 Purge the dispenser line by running several gallons of fuel into an approved safety container 12 Bleed Back Measurement Remove the dispenser authorization The turbine should stop runn...

Page 6: ... the bleed back measurement procedure and set up of the ARM Repressure Module A 3 GPH test should now be completed LEAK DETECTOR TEST AT 3 GPH Whether a new installation or the retrofit of an existing installation using VMI Leak Detectors a leak test should be performed This is done to ensure that the VMI Leak Detector is operating normally and that the selected ARM delay time is long enough for t...

Page 7: ...ZZLE position Fuel will begin to flow into the beaker After the reading on the right hand pressure gauge falls to zero set the selector to the PRESSURE STEP TEST position 8 Read Operating Pressure Authorize the dispenser The turbine will start running Watch for the turbine to achieve operating pressure as seen on the right hand pressure gauge Note this pressure reading and then remove the authoriz...

Page 8: ...Switch The turbine will start running Immediately return to the dispenser On the right hand pressure gauge of the LDT 890 verify that normal operating has been achieved Set the selector on the LDT 890 to the GPH TEST position This will introduce a 3 GPH leak to the system Because the leak detector has already stepped through to its fully open position nothing will change This leak must be introduc...

Page 9: ...3 Automatic Repressure Module is now fully functional and actively working to eliminate slow flow issues TIMER SWITCH SETTINGS Table 1 Pump Timer Switch Setting Switch Number Pump Time 1 OFF 30 seconds ON 70 seconds Table 2 Delay Timer Switch Settings Bleed Back ml Switch Number Delay Time minutes 2 3 4 5 150 OFF OFF OFF OFF 2 0 350 ON OFF OFF OFF 4 5 550 OFF ON OFF OFF 7 0 700 ON ON OFF OFF 9 5 9...

Page 10: ...om If you have any questions or concerns after reading this manual please contact VMI We are committed to customer service and the safety and satisfaction of our customers Manufactured by Vaporless Manufacturing Inc Electronics Division 8700 East Long Mesa Drive Prescott Valley Arizona 86314 Phone 928 775 5191 Website www vaporless com Technical Support electronics vaporless com ...

Reviews: